Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: Leksell Gamma Knife Perfexion, Radionuclide radiation therapy system. Article Number 715000, Elekta, Inc. Norcross, GA 30092. Teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sized os a few millimeters to several centimeters.

Recall number
Z-1230-2010
Recalling firm
Elekta, Inc.
Product
Leksell Gamma Knife Perfexion, Radionuclide radiation therapy system. Article Number 715000, Elekta, Inc. Norcross, GA 30092. Teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sized os a few millimeters to several centimeters.
Status
Terminated
Initiated
2008-11-09
Posted
2010-04-05
Terminated
2010-12-28
Reason
Radiation unit doors could close too fast on emergency exit.
Root cause
Software design
Action
Field Change Order 200 070 "MCU CS SW 8.3" dated September 11, 2008 was released. Consignees were informed of the necessary steps to avoid further issues relating to the affected product. For further information, contact Elekta, Inc. at 1-770-300-9725.
Quantity
19 units
Distribution
Nationwide Distribution -- AR, AZ, CA, FL, IL, MI, MN, MS, MO, NJ, OH, OR, PA, UT and VA.
Lot, serial or model codes
Serial Numbers: 6004, 6006, 6008, 6009, 6010, 6011, 6012, 6014, 6015, 6016, 6020, 6021, 6022, 6023, 6025, 6029, 6030, 6032 and 6036.
510(k) numbers
["K063512"]
PMA numbers
not on file
Product code
IWB
Firm FEI
1037831
Firm city
Norcross
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register