Skip to the content

FDA Medical Device Recalls (openFDA)

OriGen Biomedical, Inc.: ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV32F

Recall number
Z-1229-2019
Recalling firm
OriGen Biomedical, Inc.
Product
ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV32F
Status
Terminated
Initiated
2017-01-17
Posted
2019-06-04
Terminated
2021-07-27
Reason
Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.
Root cause
Process control
Action
On 01/17/2017, Technical Bulletin 17.01 was mailed to customers. Customers were informed to clamp all sizes of catheters only in the middle. The following contact information was provided: US: 1-512-474-7278, and Germany: +49-7475-915591.
Quantity
66
Distribution
US: NY, AB, AR, FL, MO, NC, GA, LA, CO, AL, MA, CA, MI, WI, TX, DC, OH, CT, TN, IN, OR, IL, MN, QC, SC, KY, OK, AZ, UT, RI, NJ, SD, PA, MS, NM. Foreign (OUS): UAE, Saudi Arabia, Colombia, UK, Netherlands, Italy, Czech Republic, Germany, India, Japan, Canada, Thailand, Sweden, Belgium, Spain, Poland, South Africa, Chile
Lot, serial or model codes
Lots: P18806, R19368
510(k) numbers
["K113869"]
PMA numbers
not on file
Product code
DWF
Firm FEI
1000160256
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register