Skip to the content

FDA Medical Device Recalls (openFDA)

AMS Innovative Center - San Jose: AMS brand GreenLight PV Laser System, Model-P/N: 0010-9230; Product is manufactured and distributed by AMS Solutions, San Jose, CA The GreenLight Surgical Laser System and accessories are intended for use in endoscopic (cystscopic) KTP 532 nm laser resection of the prostate for the treatment of benign prostatic hyperplasia (BPH). The entire laser unit and controls are contained in a single console that is connected to external electricity and water sources. Fibers are sterile, single-use, disposable fiber optic delivery systems, designed to deliver 532nm laser energy, the fiber is connected to the laser console and used to cut, coagulate and vaporize prostate tissue.

Recall number
Z-1229-2011
Recalling firm
AMS Innovative Center - San Jose
Product
AMS brand GreenLight PV Laser System, Model-P/N: 0010-9230; Product is manufactured and distributed by AMS Solutions, San Jose, CA The GreenLight Surgical Laser System and accessories are intended for use in endoscopic (cystscopic) KTP 532 nm laser resection of the prostate for the treatment of benign prostatic hyperplasia (BPH). The entire laser unit and controls are contained in a single console that is connected to external electricity and water sources. Fibers are sterile, single-use, disposable fiber optic delivery systems, designed to deliver 532nm laser energy, the fiber is connected to the laser console and used to cut, coagulate and vaporize prostate tissue.
Status
Terminated
Initiated
2010-02-08
Posted
2011-02-14
Terminated
2011-02-14
Reason
A failure mode whereby the fiber attachment to the GreenLight PV laser system could inadvertently be broken. If fiber were to bend sharply at point of fiber connection, it may break and cause a burn.
Root cause
Other
Action
AMS Innovative Center - San Jose issued Important Product Update Information letters via express mail on February 8, 2010, and September 15, 2010 to all its direct consignees. The letters identified the product, the problem, and the action to be taken by the customer. Customers were instructed to attach the Fiber Guard to their GreenLight PV Laser System, and to complete the Response Form in the enclosed, self-addressed, postage paid envelope to AMS. For any questions regarding this recall call the Customer Care Department at 1-800-356-7600 or send an e-mail to CustomerCare@ammd.com.
Quantity
~242,000 units
Distribution
Worldwide Distribution including US
Lot, serial or model codes
All units of the model.
510(k) numbers
["K010284"]
PMA numbers
not on file
Product code
GEX
Firm FEI
2937094
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register