FDA Medical Device Recalls (openFDA)
Thoratec Corporation: HeartMate II Left Ventricular Assist System (HM II LVAS) Sealed Outflow Graft Bend Relief. The Sealed Outflow Graft Bend Relief is provided as a component of the HM II Sealed Outflow Graft, which in turn is a component of the HM II LVAS Implant Kit with Sealed Grafts. A short configuration of the HM II LVAS Sealed Outflow Graft Bend Relief is provided as a stand-along product. Thoratec Corporation 6035 Stoneridge Drive Pleasanton, CA 94588. Product Usage: The HM II LVAS is intended to provide hemodynamic support in patients with end-stage, refractory left ventricular heart failure; either for temporary support, such as a bridge to Cardiac Transplanation or myocardial recovery, or as permanent destination therapy. The HeartMate II is intended for use inside or outside the hospital. The outflow of the HM II LVAS blood pump is connected to the aorta with a flexible vascular graft. The outflow graft is provided with a bend relief, a three or four in long tube of reinforced ePTFE surrounding the outflow graft proximal to the pump that is designed to prevent kinking of the outflow graft.
- Recall number
- Z-1228-2012
- Recalling firm
- Thoratec Corporation
- Product
- HeartMate II Left Ventricular Assist System (HM II LVAS) Sealed Outflow Graft Bend Relief. The Sealed Outflow Graft Bend Relief is provided as a component of the HM II Sealed Outflow Graft, which in turn is a component of the HM II LVAS Implant Kit with Sealed Grafts. A short configuration of the HM II LVAS Sealed Outflow Graft Bend Relief is provided as a stand-along product. Thoratec Corporation 6035 Stoneridge Drive Pleasanton, CA 94588. Product Usage: The HM II LVAS is intended to provide hemodynamic support in patients with end-stage, refractory left ventricular heart failure; either for temporary support, such as a bridge to Cardiac Transplanation or myocardial recovery, or as permanent destination therapy. The HeartMate II is intended for use inside or outside the hospital. The outflow of the HM II LVAS blood pump is connected to the aorta with a flexible vascular graft. The outflow graft is provided with a bend relief, a three or four in long tube of reinforced ePTFE surrounding the outflow graft proximal to the pump that is designed to prevent kinking of the outflow graft.
- Status
- Terminated
- Initiated
- 2012-02-23
- Posted
- 2012-03-23
- Terminated
- 2012-11-08
- Reason
- A trend of disconnected bend reliefs on the sealed outflow graft used with the HeartMate II LVAS has been identified
- Root cause
- Device Design
- Action
- Thoractec Corporation sent an Urgent Medical Device Correction letter on February 23, 2012 via overnight courier and e-mail to all affected consignees. The Urgent Medical Device Correction notification letters identify the affected products, description of the problem with illustrations, symptoms of problem and immediate actions to taken. All users are asked to review the attached labeling, including new caution statements and share with all surgical personnel. Consignees are instructed to completed and returned the Acknowledgement Form indicating that the information was received and understood. Additionally, an update concerning the Worldwide Medical Device Correction Notification is enclosed with each letter. Questions should be directed to Thoratec Regulatory Affairs at 925-847-8571.
- Quantity
- 3852 units in commerce
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of: Austria, Belgium, Denmark, Germany, Greece, France, Hong Kong, Kuwait, Ireland, Israel, Italy, Lebanon, Liechtenstein, Lithuania, Luxembourg, Malaysia, Netherlands, Turkey, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Thailand and United Kingdom.
- Lot, serial or model codes
- Sealed Outflow Graft: Catalog number 103393, all lot numbers; Sealed Outflow Short Bend Relief; Catalog number 104692, all lot numbers; Catalog number 104911, HM II LVAS Implant Kit (with sealed grafts) [US labeling] (contains catalog number 103393); all serial numbers; Catalog number 104912, HM II LVAS Implant Kit (with sealed grafts) [CE Mark labeling], contains catalog no. 103393); all serial numbers:
- 510(k) numbers
- not on file
- PMA numbers
- ["P060040","P060040S004","P060040S005"]
- Product code
- DSQ
- Firm FEI
- 2916596
- Firm city
- Pleasanton
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28