Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho Development Corporation: 10 degree Tibial Augment Wedge, a component of Ortho Balanced Knee Modular Tibial System. The device is a component of Balanced Knee Modular Tibial System and is indicated for patients suffering from severe knee pain and disability. The indications for use are: 1) loss of joint configuration and joint function, 2) Osteoarthritis of the knee joint, 3) Rheumatoid arthritis of the knee joint, 4) Post-traumatic arthritis of the knee joint, 5) Moderate valgus, varus, or flexion deformities and 6) Revision procedures where other treatments or devices have failed.

Recall number
Z-1228-2010
Recalling firm
Ortho Development Corporation
Product
10 degree Tibial Augment Wedge, a component of Ortho Balanced Knee Modular Tibial System. The device is a component of Balanced Knee Modular Tibial System and is indicated for patients suffering from severe knee pain and disability. The indications for use are: 1) loss of joint configuration and joint function, 2) Osteoarthritis of the knee joint, 3) Rheumatoid arthritis of the knee joint, 4) Post-traumatic arthritis of the knee joint, 5) Moderate valgus, varus, or flexion deformities and 6) Revision procedures where other treatments or devices have failed.
Status
Terminated
Initiated
2009-03-19
Posted
2010-04-05
Terminated
2010-06-30
Reason
A component of knee prosthesis was assembled incorrectly.
Root cause
Device Design
Action
E-mail was sent to all agents and distributors on March 19, 2009 with illustrations and instructions on how to detect the upside down assembly and how to correct the problem if detected during surgery. Direct questions to the Ortho Development Corporation by calling 1-801-619-3444.
Quantity
422 units
Distribution
United States (AZ, FL, OR, TX and UT) and Japan.
Lot, serial or model codes
Serial numbers: 567-1101; 567-1102; 567-1103; 567-1104; 567-1105; 567-1106; 567-1201; 567-1202; 567-1203; 567-1204; 567-1205 and 567-1206.
510(k) numbers
["K031201"]
PMA numbers
not on file
Product code
JWH
Firm FEI
3017298928
Firm city
Draper
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register