GE Healthcare, LLC: Replacement back up batteries distributed on or after April 1, 2019, for CARESCAPE R860, Engstrom Carestation and Engstrom PRO Ventilators.

Recall number
Z-1227-2022
Recalling firm
GE Healthcare, LLC
Product
Replacement back up batteries distributed on or after April 1, 2019, for CARESCAPE R860, Engstrom Carestation and Engstrom PRO Ventilators.
Status
Open, Classified
Initiated
2022-04-18
Posted
2022-06-06
Terminated
not on file
Reason
Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.
Root cause
Under Investigation by firm
Action
The firm, GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter the consignees via mail on 04/18/2022. The letter describes the product, problem and actions to be taken. The customers instructed to take the following actions: 1. You can continue to use the affected ventilators while the ventilator is connected to an AC mains power source that is supported by backup emergency power. 2. If it is absolutely necessary to use the ventilator by relying on the battery (such as during required transport where there are no other options), ensure you follow standard clinical practice of having a readily accessible means of appropriate alternative ventilation (for example a bag-valve system) and personnel with the capability to administer this alternative means at all times. 3. Immediately after receiving this communication, perform the Battery Performance Test as described in Appendix A. Replace the batteries when necessary, before patient use. 4. When not in patient use, it is recommended that the device always remains connected to the AC mains power source to prevent battery discharge and degradation, even when in storage. 5. It is recommended that the Battery Performance Test be completed every three months as described in Appendix A. 6. If the device has been in storage for over three months, perform the Battery Performance Test as described in Appendix A prior to use. 7. The backup batteries must be replaced at a minimum every three years. 8. Complete the attached Medical Device Notification Acknowledgement Response form and send to FMI34126RC.BATTERY@ge.co If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity
88,379 batteries
Distribution
Worldwide Distribution: US (nationwide) to AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and OUS (internationally) to Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Columbia, Cote D'Ivoire, Mexico, Uruguay, United Kingdom, United Arab Emirates, Turkey, Thailand, Tanzania, Sweden, Spain, Singapore, Saudi Arabia, Russia, Romania, Portugal, Poland, Philippines, Panama, Omar, Norway, Nigeria, Nicaragua, New Zealand, Netherlands, Malaysia, Latvia, Kuwait, Korea, Kenya, Kazakhstan, Japan, Italy, Ireland, Indonesia, India, Hungary, Germany, France, Finland, Estonia, Egypt, Ecuador, Denmark, Czech Republic.
Lot, serial or model codes
Field Replacement Unit kits PN: 1009-5682-000-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V) and PN: 5856787-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V PAIR).
510(k) numbers
["K111116","K142679"]
PMA numbers
not on file
Product code
CBK
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register