Skip to the content

FDA Medical Device Recalls (openFDA)

Covidien LP: Covidien Duet TRS Articulating 60- 3.5MM, Single Use Loading Unit Ref: DUET6035A The ENDO GIA Universal Staplers with DUET TR S Loading Units have applications in abdominal, gynecologic and pediatric surgery for resection, transection, and creation of anastomosis. They may be used for transaction and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas

Recall number
Z-1227-2012
Recalling firm
Covidien LP
Product
Covidien Duet TRS Articulating 60- 3.5MM, Single Use Loading Unit Ref: DUET6035A The ENDO GIA Universal Staplers with DUET TR S Loading Units have applications in abdominal, gynecologic and pediatric surgery for resection, transection, and creation of anastomosis. They may be used for transaction and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas
Status
Terminated
Initiated
2012-02-16
Posted
2012-03-19
Terminated
2021-03-04
Reason
Missing component results in the stapler not firing.
Root cause
Process control
Action
The firm, Covidien, sent a "Medical Device Recall" letter dated February 16, 2012 to customers via federal express. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use of Lot# N1H003LX of Duet TRS Articulating Loading Unit and segregate affected product; return inventory to Field Returns Department, 195 McDermott Road, North Haven, CT 06473 and contact customer service at SDFeedback@Covidien.com or (800) 722-8772, option #1 for RGA # prior to returning; and complete and return the Recall Product Return Form via fax to (203) 822-6009 (all affected units must be returned with completed form through the Distributor). If you have any questions or concerns, please do not hesitate to contact your Covidien representative or Quality Assurance at 203-492-5232, or Customer Service at 800-722-877 option 1 customer service.
Quantity
1158 units
Distribution
Worldwide distribution: USA (nationwide) and country of: Venezuela.
Lot, serial or model codes
Lot Number: N1H0037LX
510(k) numbers
["K080898"]
PMA numbers
not on file
Product code
GDW
Firm FEI
1219930
Firm city
North Haven
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register