Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Xomed, Inc.: NIM Vital Patient Interface 4.0 (P/N: NIM4CPB1) NIM with Software version 1.3.2

Recall number
Z-1224-2022
Recalling firm
Medtronic Xomed, Inc.
Product
NIM Vital Patient Interface 4.0 (P/N: NIM4CPB1) NIM with Software version 1.3.2
Status
Open, Classified
Initiated
2022-05-03
Posted
2022-06-02
Terminated
not on file
Reason
Software anomaly was identified.
Root cause
Software Design Change
Action
The firm issued a recall notification to consignees on 05/03/2022 via mail. The letter instructs the consignee to follow the IFU regarding the setup and "Lead Off Detected" error message, forward the notification to those who need to be aware within the organization and to any location which the devices have been distributed, and to complete the response form. An additional correction notification was sent on about 09/13/2023 to notify consignees of the update availability (NIM Vital" SW version v1.4.3) to fix this issue.
Quantity
29 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of TX, CO, NJ and the country of France.
Lot, serial or model codes
Model Number: NIM4CPB1; UDI/DI: 00763000002985, lot numbers: P1910256, P1910260, P1910402, P1910404, P2008073, P2008130, P2008131 and 00763000395902, lot numbers: P2026275, P2026290, P2026329, P2026330, P2026357, P2026388, P2026389, P2026680, P2026681, P2026718, P2026719, P2026793, P2026956, P2026957, P2026958, P2026959, P2026981, P2026994, P2026995, P2127005, P2127006, P2127013
510(k) numbers
["K200759"]
PMA numbers
not on file
Product code
GWF
Firm FEI
1045254
Firm city
Jacksonville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register