FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Baxter Colleague CX Single Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A., Made in Singapore; product code 2M8161

Recall number
Z-1224-2007
Recalling firm
Baxter Healthcare Corp.
Product
Baxter Colleague CX Single Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A., Made in Singapore; product code 2M8161
Status
Terminated
Initiated
2007-08-15
Posted
2007-09-19
Terminated
2008-02-21
Reason
Some of the pumps were upgraded by one service technician who failed to replace prisms and batteries as required in the upgrade.
Root cause
Other
Action
The affected hospitals were sent an Urgent Colleague Deployment Correction letter dated 8/15/07 via overnight mail. The hospitals were informed that some of the Colleague pumps at their facility may not have been completely upgraded to include all the necessary components required as part of Baxter's Corrective Action Plan (CAP). The hospitals were provided a list of affected pump serial numbers and were requested to remove the pumps from clinical use and notify all end users of the action. The accounts were requested to complete the enclosed customer reply form, acknowledging the receipt and understanding of the letter and fax it back to Baxter at 1-847-270-5457. The hospitals were also requested to report any malfunction or adverse event related to the Colleague Infusion Pumps to Baxter Corporate Product Surveillance at 1-800-437-5176.
Quantity
42 pumps
Distribution
Florida and California
Lot, serial or model codes
product code 2M8161, serial numbers 14051565CC, 15034273CC, 15034492CC, 15040255CC, 15041116CC, 15041169CC, 15041170CC, 15041172CC, 15041181CC, 15041194CC, 15041199CC, 15041206CC, 15041212CC, 15041218CC, 15041241CC, 15041288CC, 15041290CC, 15041317CC, 15050020CC, 15050034CC, 15050045CC, 15050061CC, 15050080CC, 15050117CC, 15050129CC, 15060087CC, 15060217CC, 15060253CC, 15060479CC, 15060489CC, 15060549CC, 15060557CC, 15060600CC, 15060738CC, 15060790CC, 15060827CC, 15060939CC, 15060940CC, 15060968CC, 15063211CC, 15075371CC, 15075103CC
510(k) numbers
["K063696"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register