FDA Medical Device Recalls (openFDA)
Elekta, Inc.: Leksell GammaPlan, Model 5.34. Intended to be used for planning the dosimetry of treatments in stereotactic radiation therapy.
- Recall number
- Z-1223-2011
- Recalling firm
- Elekta, Inc.
- Product
- Leksell GammaPlan, Model 5.34. Intended to be used for planning the dosimetry of treatments in stereotactic radiation therapy.
- Status
- Terminated
- Initiated
- 2005-10-01
- Posted
- 2011-02-14
- Terminated
- 2011-03-02
- Reason
- Investigation found that if a user accidentally selects the wrong image/tube position images can be displayed flipped.
- Root cause
- Other
- Action
- Product bulletin Leksell GammaPlan - "Leksell GammaPlan version 5.34" (dated October 2005) was sent to affected Customers as notification of the problem. Field Change Order FCO-DCO-05-0006 (dated November 22, 2005) and Technical Notes Leksell GammaPlan Leksell SurgiPlan Important Notice - "Potential issue using flipped angiographic images" (dated October 2005) was sent to Elekta Service Engineers as notification of the problem. The letters described the issue and the FCO asked the Service Engineers to confirm their receipt of the documentation. The Technical Notes also discussed image visualization requirements and listed recommendations.
- Quantity
- 272 units
- Distribution
- Worldwide Distribution
- Lot, serial or model codes
- SG33620363, US68537171, SG10821625, 3713G01281, SG03321319, SG33520213, US19502892, C160, SG0320704, SG44120295, 00000000000002, SG31630678, SG24820212, SG24820194, SG44221038, SG34620630, SG40920412, SG31020878, A4986A, SG11520160, SG44720158, SG41720310, SG43820213, 3416G09539, SG24820139, US19257221, SG33820630, 19476882, SG13320399H, SG40520126, SG50425261, SG41720317, US66320311, US19490144, SG21220130, US68154038, US19501776, SW ONLY, SG32520836, 3416G08464, SG21220113, SG44925638, SG04120289, SG51220179, SG2030679, US6824174, SG22920793, SG4820232, 3628G06513, 000000000000019, 00000000000004, US19310946, 3416G05185, US69131975, SG42520530, SG23320693, US68104248, SG23720396, SG24820231, SG14720637, SG43520759, US68513855, SG44820136, SG20220701, SG44120292, US19476874, US6863578, SG04020719, 3416G07810, SG43720660, SG44925635, US68538435, US19322820, SG33320960, US19319754, SG31620624, US10102906, SG44925637, SG43520477, US69172955, SG42720416, 3711C04231, US68154724, SG0620611, SG20220192, SG33620337, US19311023, SG35020182, US19239028, SG50425266, SG51920134, 000000000000011, SG44925633, SG24021335, 0304011939, A4986A, SG34220901, SG1220108, SG32120124, US69132547, SG43720550, US19476879, 6146A06124, SG41920535, 6412A00007, US68500234, US19322145, SG20220694, US68115508, SG41620718, SG10720353, 000000000000014, SG43520485, SG42820297, 000000000000015, SG41020267, US68104599, IE24488098, SG11421969, SG41720337, US6836799, US19344670, SG11720734, US68064889, SG20220679, 000000000000016, SG20220699, SG11421978, SG34220707, US19378673, SG34420200, SG20220143, SG50425271, US10025765, US90345020, US19345857, and SG13420315.
- 510(k) numbers
- ["K973441"]
- PMA numbers
- not on file
- Product code
- IWB
- Firm FEI
- 1037831
- Firm city
- Norcross
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28