Skip to the content

FDA Medical Device Recalls (openFDA)

K C Pharmaceuticals Inc: Top Care Sterile Preserved Saline Solution 12oz, Product SKU#0036800-16040. This product is supplied in 12 fl. oz. (355 mL) bottles for use in rinsing of soft (hydrophilic) contact lenses after heat (thermal) disinfection and rinsing of contact lenses following chemical disinfection.

Recall number
Z-1223-2010
Recalling firm
K C Pharmaceuticals Inc
Product
Top Care Sterile Preserved Saline Solution 12oz, Product SKU#0036800-16040. This product is supplied in 12 fl. oz. (355 mL) bottles for use in rinsing of soft (hydrophilic) contact lenses after heat (thermal) disinfection and rinsing of contact lenses following chemical disinfection.
Status
Terminated
Initiated
2010-02-03
Posted
2010-04-02
Terminated
2011-01-05
Reason
The recall was initiated because the affected lots were manufactured between two failed media fills; therefore the lots produced cannot be assured as being free of contamination and may not be sterile.
Root cause
Other
Action
An "Urgent: Medical Device Recall Notification" letter dated February 17, 2010 was issued to customers. Consignees were instructed to immediately hold and discontinue selling the existing, affected stock. Consignees were also instructed to check inventories at all distribution locations for affected products shipped between June 8, 2OO9 and August 7, 2009 and perform a SUBRECALL to the retail level and report back to KC Pharmaceuticals. If the retailers indicate they have any of the affected lots of product, please have them contact KC Pharmaceuticals at 1-888-527-4276 for a shipping authorization number. Consignees were asked to complete the enclosed Recall Effectiveness Check Form, even if they did not have the products in question and return the form to KC Pharmaceuticals via mail within five (5) business days. The recalled lot and completed packing slips were asked to be shipped to: KC Pharmaceuticals-SALP Recall, 3201 Producer Way, Pomona, CA 91768. Direct questions to KC Pharmaceuticals by calling 1-888-527-4276, extension 27.
Quantity
11,052 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
Lot #9D524.
510(k) numbers
not on file
PMA numbers
["P930005"]
Product code
LYL
Firm FEI
2026940
Firm city
Pomona
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register