Skip to the content

FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: Brand Name: Single Use Electrosurgical Knife Product Name: Olympus Single Use Electrosurgical Knife KD-640L Model/Catalog Number: KD640-L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.

Recall number
Z-1222-2026
Recalling firm
Olympus Corporation of the Americas
Product
Brand Name: Single Use Electrosurgical Knife Product Name: Olympus Single Use Electrosurgical Knife KD-640L Model/Catalog Number: KD640-L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.
Status
Open, Classified
Initiated
2024-02-17
Posted
2026-01-30
Terminated
not on file
Reason
A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.
Root cause
Device Design
Action
On December 17, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Carefully read the content of this notification. 2. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification, including the instruction for intermittent output activation and applicable Warnings and Cautions. 3. Keep a copy of this notification with the Instructions for Use for any affected devices remaining in your inventory. 4. If you have further distributed the affected product, please identify your customers and forward this notification to them. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail Cynthia.Ow@Olympus.com.
Quantity
units
Distribution
US Nationwide distribution.
Lot, serial or model codes
Model/Catalog Number: KD640-L UDI: 04953170208423 Lot Number(s): 2ZK, 31K, 32K, 33K, 34K, 35K, 36K, 37K, 38K, 39K, 3XK, 3YK, 3ZK, 41K, 42K, 43K, 44K, 45K, 46K, 47K, 48K, 49K, 4XK, 4YK, 4ZK, 51K, 52K, 53K, 54K, 55K, 58K, 59K
510(k) numbers
["K092309"]
PMA numbers
not on file
Product code
KNS
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register