Skip to the content

FDA Medical Device Recalls (openFDA)

Tyber Medical: A.L.P.S. mvX-MEDIAL TIBIA PLATE 10H RT-Medial Tibia Plate, 10-Hole, Right Intended Fixation of fractures of the distal tibia. Model/Catalog Number: 770715102

Recall number
Z-1222-2025
Recalling firm
Tyber Medical
Product
A.L.P.S. mvX-MEDIAL TIBIA PLATE 10H RT-Medial Tibia Plate, 10-Hole, Right Intended Fixation of fractures of the distal tibia. Model/Catalog Number: 770715102
Status
Open, Classified
Initiated
2025-01-21
Posted
2025-02-21
Terminated
not on file
Reason
Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Root cause
Process control
Action
Tyber Medical issued Urgent Medical Device Recall letter on 1/24/25 via email to the Distributor. Letter states reason for recall, health risk and action to take: 1. Acknowledge the receipt of this communication via email correspondence to feedback@tybermed.com. 2. Further distribute this communication to all applicable parties and/or customers that possess the impacted inventory. 3. Identify via laser-marked lot number and quarantine the impacted plates under Zimmer Biomet s control that have not been distributed for sale. An RMA will be provided for the return of goods to Tyber Medical for further inspection. 4. Identify via laser-marked lot number and quarantine the impacted plates that have been distributed. An RMA will be provided for the return of goods to Tyber Medical for further inspection. 5. Complete the customer response form located in Appendix 2 and return it to Tyber Medical via email correspondence to feedback@tybermed.com.
Quantity
209 units
Distribution
Nationwide
Lot, serial or model codes
UDI-DI 00196449016006; Lot Numbers 265815, 24055DF03, 24055DF04
510(k) numbers
["K232652"]
PMA numbers
not on file
Product code
HRS
Firm FEI
3012966183
Firm city
Bethlehem
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register