Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho-Clinical Diagnostics: VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides, VITROS Chemistry Products MicroTip Reagents, and VITROS Immunodiagnostic Products Reagents

Recall number
Z-1222-2019
Recalling firm
Ortho-Clinical Diagnostics
Product
VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides, VITROS Chemistry Products MicroTip Reagents, and VITROS Immunodiagnostic Products Reagents
Status
Terminated
Initiated
2019-03-11
Posted
not on file
Terminated
2020-12-07
Reason
One of the software algorithms used to detect sample dispense errors was inadvertently disabled. Because of this, sample dispense errors may lead to incorrect results being reported without an error code to alert the user.
Root cause
Software design
Action
Urgent Product Correction Notification letters dated 3/11/19 were sent to customers.
Quantity
80
Distribution
Worldwide Distribution - US Nationwide in the states of: AZ, CA, FL, IL, OH, TX, and VA. The products were distributed to the following foreign countries: Australia, Belgium, Canada, Chile, Columbia, France, India, Italy, Japan, Portugal, Singapore, Spain, United Kingdom
Lot, serial or model codes
Systems Running Software Version 3.4 & 3.4.1
510(k) numbers
["K182063"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1319681
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register