Skip to the content

FDA Medical Device Recalls (openFDA)

C.R. Bard Inc: The Bard Intra-abdominal Pressure Monitoring Device is composed of a tubing set used for infusing fluid into the urinary bladder through the Foley catheter sampling port. It utilizes a clamping device to occlude the urinary drainage tubing to form a fluid column through which pressure is measured.

Recall number
Z-1219-2020
Recalling firm
C.R. Bard Inc
Product
The Bard Intra-abdominal Pressure Monitoring Device is composed of a tubing set used for infusing fluid into the urinary bladder through the Foley catheter sampling port. It utilizes a clamping device to occlude the urinary drainage tubing to form a fluid column through which pressure is measured.
Status
Terminated
Initiated
2020-01-13
Posted
not on file
Terminated
2021-04-12
Reason
Leaks or disconnect near the sampling port of the Intra Abdominal Pressure Monitoring Device which could lead to the potential of inaccurate readings. The measurement should be confirmed with other means, such as a new IAP system of the same kind or by choosing other diagnostic means such as but not limited to alternative systems, such as open IAP measurement options or exchanging the tubing system, which does not require invasive maneuvers.
Root cause
Device Design
Action
Letters were sent 01/13/2020 to the user level instructing the return of two specific product lots of IAP001, Lots NGDQ2943 and NGDS0622.
Quantity
1,300 total
Distribution
Distributed Nationwide including Washington DC and Guaynabo PR; Distributed OUS to Belgium and Japan.
Lot, serial or model codes
Product Code IAP001 lot numbers NGDQ2943 and NGDS0622.
510(k) numbers
["K070201"]
PMA numbers
not on file
Product code
PHU
Firm FEI
1018233
Firm city
Covington
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28