FDA Medical Device Recalls (openFDA)

Abbott Laboratories HPD/ADD/GPRD: AxSYM Homocysteine Reagent Pack, list 5F51-20, Produced by Axis-Shield, Oslo, Norway, for Abbott Laboratories, Abbott Park, IL 60064 USA

Recall number
Z-1218-03
Recalling firm
Abbott Laboratories HPD/ADD/GPRD
Product
AxSYM Homocysteine Reagent Pack, list 5F51-20, Produced by Axis-Shield, Oslo, Norway, for Abbott Laboratories, Abbott Park, IL 60064 USA
Status
Terminated
Initiated
2003-08-22
Posted
2003-09-11
Terminated
2004-07-20
Reason
There is a kit-to-kit difference within the lots, with the reagents generating different polarization values than those in others.
Root cause
Other
Action
Recall letters dated 8/22/03 informed users of the kit to kit difference within the lots, and outlined the actions needed to take for patient results obtained from the use of the affected kits. The customers were requested to destroy any remaining inventory of the affected reagents. Any questions were directed to Abbott's Customer Service at 1-877-422-2688.
Quantity
3505 kits
Distribution
Lot 01724M300 was distributed nationwide, as well as to Canada, Singapore, Tahiland, Korea, Hong Kong and China. Lots 03314HN00 and 03533HN00 were distributed by Abbott's Delkenheim, Germany subsidiary worldwide outside of the United States. These two lots were not distributed in the U.S.
Lot, serial or model codes
list 5F51-20, lots 01724M300, 03314HN00, 03533HN00
510(k) numbers
["K992858"]
PMA numbers
not on file
Product code
LPS
Firm FEI
1415939
Firm city
Abbott Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register