Skip to the content

FDA Medical Device Recalls (openFDA)

Exactech, Inc.: Gibralt Spine System Polyaxial Screw Intended to promote fusion of the cervical spine, and the thoracic spine, (C3-T3), the Gibralt Spinal System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C3-T3) spine.

Recall number
Z-1217-2014
Recalling firm
Exactech, Inc.
Product
Gibralt Spine System Polyaxial Screw Intended to promote fusion of the cervical spine, and the thoracic spine, (C3-T3), the Gibralt Spinal System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C3-T3) spine.
Status
Terminated
Initiated
2014-01-06
Posted
2014-03-18
Terminated
2014-08-04
Reason
The products have been identified to contain springs which have been manufactured with type-304 stainless steel rather than the specified titanium alloy.
Root cause
Mixed-up of materials/components
Action
Exactech sent an "URGENT PRODUCT RECALL NOTICE" dated January 6, 2014, to all affected customers. The letter identified the product and action needed to taken by the customers. Customers were instructed to: 1) Immediately cease distribution. 2) Extend the information to accounts in possession of the product. 3) Verify if they have any of the subject Gibralt Screws from the specified lots. 4) Complete and fax back the attached form. 5) Return affected products. For further questions, please call ( 352) 327-4847.
Quantity
242
Distribution
USA Distribution including the states of FL, NY, OR, IL, PA, MO and NJ.
Lot, serial or model codes
size 3.5mm x 10mm, Catalog #05-000-20-3510, Lot Number #35917001. size 3.5mm x 14mm, Catalog #05-000-20-3514, Lot Number 35917002 and 40023002. size 3.5mm x 20mm, Catalog #05-000-20-3520, Lot Number 35917003. size 3.5mm x 26mm, Catalog #05-000-20-3526, Lot Number 35917004. size 4.0mm x 14mm, Catalog #05-000-20-4014, Lot Number 35917005. size 4.0mm x 20mm, Catalog #05-000-20-4020, Lot Number 35917006. size 4.0mm x 26mm, Catalog #05-000-20-4026, Lot Number 35917007. 1/4 Shank, size 3.5mm x 26mm, Catalog #05-000-20-3526, Lot Number 35917008. 1/4 Shank, size 3.5mm x 30mm, Catalog #05-000-20-3530, Lot Number 35917009. 1/4 Shank, size 4.0mm x 26mm, Catalog #05-000-20-4026, Lot Number 35917010. 1/4 Shank, size 4.0mm x 30mm, Catalog #05-000-20-4030, Lot Number 35917011.
510(k) numbers
["K110197"]
PMA numbers
not on file
Product code
MNI
Firm FEI
1038671
Firm city
Gainesville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register