Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Healthcare Inc.: Philips EasyDiagnost Eleva DRF, Fluoroscopic x-ray system Software Release 3.0 and 3.1 Model: 706034 Product Usage: Radiological image processing system (computed radiography x-ray system)

Recall number
Z-1216-2012
Recalling firm
Philips Healthcare Inc.
Product
Philips EasyDiagnost Eleva DRF, Fluoroscopic x-ray system Software Release 3.0 and 3.1 Model: 706034 Product Usage: Radiological image processing system (computed radiography x-ray system)
Status
Terminated
Initiated
2012-02-14
Posted
2012-03-15
Terminated
2020-09-23
Reason
Misinterpretation of Mirror Icon-Mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system
Root cause
Software design
Action
Philips Healthcare sent an Urgent Medical Device Correction letter dated February 8, 2012 to all affected customers. The letter identified the affected product, problem description, hazard involved, actions to be taken by customer/user and actions planned by Philips. The letter contains important information for continue safe and proper use of the equipment. Philips will contact customers to schedule the installation of the updated software. For further information or support contact your local Philips representative at the Technical Support Line 866-767-2822.
Quantity
60 units
Distribution
Worldwide Distribution - USA Nationwide
Lot, serial or model codes
DRF systems with software reI 3.0 and 3.1 S/N: 362600 362601 364209 364210 375245 375850 376439 416254/SN09000007 417890/SN08000120 384396 401281/SN08000142 399440 387492 399621 398185 423047 399623 422094 422823 412477/SN08000137 425981 402409 402410/SN08000112 428620 428623 426466 426467 404548/SN08000110 417116/SN08000126 410563/SN08000120 408593/SN08000103 408592/SN08000111 408237/SN08000115 409087/SN08000107 435412 413514/SN08000123 413831/SN08000122 422648 422650 422652 414463 416393/SN08000140 416643/SN09000005 420212 421780 420872/SN09000003 425745 425744 427503 435926 430916 H706034090029 429498 431820 435419 439922 432277 435422
510(k) numbers
["K031535"]
PMA numbers
not on file
Product code
KPR
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register