Skip to the content

FDA Medical Device Recalls (openFDA)

Konica Minolta Medical Imaging USA, Inc.: Sonimage HS1 Ultrasound Kit AC adapter, Product Number: A8AR, UDI: 04560141941588

Recall number
Z-1215-2018
Recalling firm
Konica Minolta Medical Imaging USA, Inc.
Product
Sonimage HS1 Ultrasound Kit AC adapter, Product Number: A8AR, UDI: 04560141941588
Status
Terminated
Initiated
2017-12-04
Posted
not on file
Terminated
2019-10-28
Reason
There is a risk of circuit board breakage within the main housing of the AC Adapter unit. Smoke may be emitted from the AC Adapter as a direct result.
Root cause
Device Design
Action
Konica Minolta is preparing for the final AC Adapter replacement. Upon receipt of the new AC Adapter, owners of SONIMAGE HSl System will be asked to replace the existing AC Adapter. Additionally, instructions for the return of the current AC Adapter will be provided.
Quantity
17
Distribution
Distributed in the following U.S. states: AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, WA, WI and the District of Columbia. Distributed in Canada and Brazil.
Lot, serial or model codes
Serial Numbers (each has "A8AR-" prefix): 00068 00031 00062 00066 00063 00047 00065 00032 00064 00028 00069 00060 00027 00067 00061 00001 00002
510(k) numbers
["K151060"]
PMA numbers
not on file
Product code
IYN
Firm FEI
2241281
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register