FDA Medical Device Recalls (openFDA)
Konica Minolta Medical Imaging USA, Inc.: Sonimage HS1 Ultrasound Kit AC adapter, Product Number: 1. A9RP, UDI: 04560141945463; 2. AAEU, UDI: 04560141946385; 3. A9RR, UDI: 04560141945470
- Recall number
- Z-1214-2018
- Recalling firm
- Konica Minolta Medical Imaging USA, Inc.
- Product
- Sonimage HS1 Ultrasound Kit AC adapter, Product Number: 1. A9RP, UDI: 04560141945463; 2. AAEU, UDI: 04560141946385; 3. A9RR, UDI: 04560141945470
- Status
- Terminated
- Initiated
- 2017-12-04
- Posted
- not on file
- Terminated
- 2019-10-28
- Reason
- There is a risk of circuit board breakage within the main housing of the AC Adapter unit. Smoke may be emitted from the AC Adapter as a direct result.
- Root cause
- Device Design
- Action
- Konica Minolta is preparing for the final AC Adapter replacement. Upon receipt of the new AC Adapter, owners of SONIMAGE HSl System will be asked to replace the existing AC Adapter. Additionally, instructions for the return of the current AC Adapter will be provided.
- Quantity
- 30
- Distribution
- Distributed in the following U.S. states: AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, WA, WI and the District of Columbia. Distributed in Canada and Brazil.
- Lot, serial or model codes
- A9RP Serial Numbers (each has "A9RP-" prefix): 00002 00021 00022 00017 00016 00080 00029 00030 00019 00023 00008 00024 00079 00020 00027 00003 00025 00028 00026; AAEU Serial Numbers (each has "AAEU-" prefix):00001 00016 00017 00018 00019 00020 00021 00022 00023 00024 00025
- 510(k) numbers
- ["K162065"]
- PMA numbers
- not on file
- Product code
- IYN
- Firm FEI
- 2241281
- Firm city
- Wayne
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28