Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: Baxter MiniCap Disconnect Cap with Povidone-Iodine Solution, 5C4466P; Accessory to disposable administration set for peritoneal dialysis

Recall number
Z-1213-2023
Recalling firm
Baxter Healthcare Corporation
Product
Baxter MiniCap Disconnect Cap with Povidone-Iodine Solution, 5C4466P; Accessory to disposable administration set for peritoneal dialysis
Status
Open, Classified
Initiated
2023-02-01
Posted
2023-03-03
Terminated
not on file
Reason
MiniCap Disconnect Cap with Povidone-Iodine are packaged in foil pouches, which may have been incorrectly sealed, i.e., the pouches may have open or weak seals. This could lead to exposure to air, resulting in insufficient iodine/dry sponge inside the MiniCap, which could lead to the potential for inadequate disinfectant.
Root cause
Process control
Action
Baxter Healthcare initiated the recall by issuing an Urgent Medical Device Recall notice to its consignees by letter on 02/01/2023. The notice requested the following from customers: 1. Check your stock and set aside all product with the affected lot numbers. 2. Contact Baxter HomeCare Services to arrange for return of impacted product and for ordering replacement product. 3. If all of your product is impacted by this recall and you are unable to get unaffected lot numbers, resulting in a significant delay in therapy, please follow the below steps: a. Contact your clinic for clinical guidance and check each pouch and do not use the MiniCap contained inside of any pouches that appear to have damaged seals. For general questions regarding this communication, contact Baxter Corporate Product Surveillance at 800-437-5176, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday." Recall was expanded and Baxter Healthcare sent an updated notification to customers and patients on 05/09/2023. The letter instructed consignees and patients to immediately cease use of the affected product, arrange for the return of affected product and assess each patient's individual next steps based on clinical evaluation. Customers were also instructed to complete the customer response form via the portal (for products directly received from Baxter) or through the distributor. If the products were further distributed, please notify customers.
Quantity
33,866,400 units
Distribution
US Nationwide distribution and Canada.
Lot, serial or model codes
UDI/DI 00085412007694, Lot numbers: GD912051, exp 31-Mar-2024; GD912068, exp 30-Apr-2024; GD912099, exp 30-Apr-2024; GD912112, exp 30-Apr-2024; GD912143, exp 30-Apr-2024; GD912204, exp 31-May-2024. Expansion on 05/10/2023: Product Code 5C4466P with all lots numbers manufactured prior to 09/30/2022.
510(k) numbers
["K152129"]
PMA numbers
not on file
Product code
KDJ
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register