Skip to the content

FDA Medical Device Recalls (openFDA)

Mar Cor Purification: Mar Cor Purification Millenium HX Portable Reverse Osmosis System. Model 3025975. Product Usage: Water Purification System is intended to be used as a dialysis accessory to produce water through reverse osmosis for use with hemodialysis equipment..

Recall number
Z-1213-2014
Recalling firm
Mar Cor Purification
Product
Mar Cor Purification Millenium HX Portable Reverse Osmosis System. Model 3025975. Product Usage: Water Purification System is intended to be used as a dialysis accessory to produce water through reverse osmosis for use with hemodialysis equipment..
Status
Terminated
Initiated
2013-06-26
Posted
2014-03-14
Terminated
2015-01-22
Reason
High inlet water pressure beyond specification and the solenoid valve does not meet application which could result in a delay in treatment.
Root cause
Component design/selection
Action
Mar Cor sent an Urgent Medical Device Correction letter dated March 5, 2014 to all affected customers. The letter identifies the affected devices, what the issue is and tells customers to contact Mar Cor to schedule correction. A response form was asked to be completed and returned. For questions contact Scott Nichols, Project Coordinator/Service Trainer at 1-800-633-3080.
Quantity
232 units
Distribution
USA Nationwide Distribution including Puerto Rico.
Lot, serial or model codes
Serial Numbers: 1294627, 1298383, 1298386, 1299352, 1299353, 1299354, 1299355, 1299356, 1299486, 1299487, 1299488, 1299489, 1299503, 1299572, 1299712, 1299713, 1299716, 1299718, 1299719, 1299720, 1299740, 1299742, 1299777, 1299779, 1299780, 1299831, 1299835, 1299850, 1299867, 1299868, 1299895, 1299897, 1299992, 1299993, 1299994, 1299996, 1299997, 1300138, 1300224, 1300248, 1300363, 1300364, 1300367, 1300368, 1300558, 1300560, 1300561, 1300562, 1300564, 1300565, 1300621, 1300649, 1300650, 1300700, 1300701, 1301081, 1301116, 1301360, 1301361, 1301386, 1301888, 1301990, 1301992, 1301997, 1302030, 1302032, 1302149, 1302214, 1302226, 1302342, 1302345, 1302346, 1302385, 1302430, 1302447, 1302448, 1302452, 1302456, 1302464, 1302517, 1302555, 1302558, 1302642, 1302696, 1302706, 1302708, 1302748, 1302767, 1302768, 1302810, 1302813, 1302922, 1302923, 1302924, 1302925, 1302926, 1302927, 1302987, 1303005, 1303108, 1303315, 1303316, 1303346, 1303348, 1303351, 1303352, 1303505, 1303516, 1303532, 1303533, 1303534, 1303557, 1303558, 1303625, 1303626, 1303689, 1303690, 1303693, 1303702, 1303711, 1303768, 1303781, 1303799, 1303806, 1303821, 1303872, 1303877, 1303908, 1303920, 1303975, 1303976, 1304070, 1304076, 1304077, 1304081, 1304163, 1304166, 1304167, 1304170, 1304173, 1304174, 1304194, 1304196, 1304203, 1304273, 1304293, 1304294, 1304297, 1304333, 1304340, 1304343, 1304344, 1304346, 1304349, 1304376, 1304381, 1304501, 1304502, 1304503, 1304507, 1304508, 1304551, 1304585, 1304587, 1304724, 1304725, 1304775, 1304778, 1304779, 1304829, 1304870, 1304902, 1304903, 1304929, 1305006, 1305007, 1305038, 1305040, 1305041, 1305087, 1305088, 1305089, 1305131, 1305133, 1305210, 1305238, 1305239, 1305355, 1305363, 1305608, 1305609, 1305650, 1305651, 1305655, 1305698, 1305700, 1305701, 1305717, 1305732, 1305735, 1305750, 1305751, 1305752, 1305795, 1305928, 1305931, 1305932, 1305975, 1305995, 1306007, 1306042, 1306046, 1306078, 1306125, 1306154, 1306178, 1306209, 1306215, 1306226, 1306259, 1306322, 1306341, 1306428, 1306429, 1306430, 1306431, 1306432, 1306654, 1306662, 1306663, 1306664, 1306665.
510(k) numbers
["K110578"]
PMA numbers
not on file
Product code
FIP
Firm FEI
3006475063
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register