FDA Medical Device Recalls (openFDA)
Konica Minolta Medical Imaging USA, Inc.: Sonimage HS1 Ultrasound Kit AC adapter, Product Number: A7AP, UDI: 04560141940031
- Recall number
- Z-1212-2018
- Recalling firm
- Konica Minolta Medical Imaging USA, Inc.
- Product
- Sonimage HS1 Ultrasound Kit AC adapter, Product Number: A7AP, UDI: 04560141940031
- Status
- Terminated
- Initiated
- 2017-12-04
- Posted
- not on file
- Terminated
- 2019-10-28
- Reason
- There is a risk of circuit board breakage within the main housing of the AC Adapter unit. Smoke may be emitted from the AC Adapter as a direct result.
- Root cause
- Device Design
- Action
- Konica Minolta is preparing for the final AC Adapter replacement. Upon receipt of the new AC Adapter, owners of SONIMAGE HSl System will be asked to replace the existing AC Adapter. Additionally, instructions for the return of the current AC Adapter will be provided.
- Quantity
- 88
- Distribution
- Distributed in the following U.S. states: AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, WA, WI and the District of Columbia. Distributed in Canada and Brazil.
- Lot, serial or model codes
- Serial Numbers (each has "A7AP-" prefix): 00045 00041 00004 00076 00090 00016 00037 00038 00060 00059 00091 00028 00089 00080 00077 00039 00057 00040 00007 00043 00024 00071 00003 00064 00011 00052 00027 00006 00044 00042 00063 00078 00087 00047 00031 00048 00008 00068 00095 00103 00096 00098 00097 00099 00100 00101 00102 00082 00084 00092 00083 00069 00079 00061 00019 00072 00081 00051 00018 00017 00065 00055 00053 00012 00015 00073 00023 00074 00002 00075 00066 00094 00020 00070 00056 00049 00025 00085 00093 00046 00086 00022 00009 00010 00021 00013 00014 00029
- 510(k) numbers
- ["K142197"]
- PMA numbers
- not on file
- Product code
- IYN
- Firm FEI
- 2241281
- Firm city
- Wayne
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28