FDA Medical Device Recalls (openFDA)
Linet Spol. S.r.o.: Sprint 200 wheeled stretcher, Item numbers 1ES200-7, 1ES210-1, and 1ES211-1.
- Recall number
- Z-1211-2023
- Recalling firm
- Linet Spol. S.r.o.
- Product
- Sprint 200 wheeled stretcher, Item numbers 1ES200-7, 1ES210-1, and 1ES211-1.
- Status
- Open, Classified
- Initiated
- 2023-01-18
- Posted
- 2023-03-04
- Terminated
- not on file
- Reason
- Labeling being updated to assure user is securing the siderail to a locked position to prevent potential patient fall and injury.
- Root cause
- Labeling design
- Action
- The firm issued "URGENT MEDICAL DEVICE CORRECTION" recall letters dated 1/17/2023 via email on 1/18/2023. The letter explains the issue that an audible clicking sound can be heard when the side rail is raised, but this sound does not indicate the side rail has locked into the raised position. The consignee is to take immediate action to notify caregivers using the Sprint 200 of this potential hazard and proper use for securing a side rail in the raised position. The consignee is informed that Linet will be adding additional safety warning labels located at each side rail latch (2 per side of stretcher), add additional safety warning labels to both side rails (1 per side of stretcher), and that additional training material and an updated copy of the Sprint 200 Emergency Stretcher Instructions for Use and Technical Description manual is available using the QR code in the letter or by following a link to access the materials. The firm's service department will be contacting the consignees to schedule an appointment to perform the corrections. Photos of the additional warning labels are included in the letter. Please contact LINET service department at 877-815-9897 with immediate questions or concerns.
- Quantity
- 335 stretchers (310 US, 25 OUS)
- Distribution
- US Distribution was made to AL, CA, FL, IA, KY, LA, MN, NV, NC, OK, TX, VA, WA, and WV. There was no government/military distribution. Foreign distribution was made to Australia, Bahrain, Belgium, Germany, Great Britain, Italy, Netherland, Singapore, South Africa, Switzerland, and United Emirates Canada.
- Lot, serial or model codes
- 1ES200-7 - Serial numbers 20220203852, 20220203853, 20220203854, 20220203407, 20220203429, 20220203425, 20220203427, 20220203428, 20220203430, 20220203836, 20220203864, 20220203402, 20220203404, 20220203420, 20220203422, 20220203841, 20220203850, 20220203858, 20220203862, 20220203865; UDI-DI 08592654358467. 1ES210-1 - Serial numbers 20220204642, 20220204649, 20220204654, 20220204666, 20220207624, 20220261655, 20220261670, 20220261671, 20220261672, 20220204637, 20220204638, 20220204639, 20220204641, 20220204646, 20220204651, 20220204656, 20220204657, 20220207626, 20220207630, 20220261659, 20220261660, 20220261650, 20220261661, 20220261662, 20220261663, 20220261664, 20220261665, 20220261666, 20220261667, 20220261668, 20220261669, 20220261676, 20220204665, 20220207622, 20220207631, 20220207634, 20220207637, 20220207638, 20220207641, 20220207642, 20220207644, 20220207645, 20220261648, 20220261649, 20220261651, 20220261656, 20220261673, 20220261674, 20220261677, 20220261678, 20220261679, 20220261680, 20220261681, 20220261682, 20220261683, 20220261657, 20220261675, 20220190550, 20220190552, 20220190554, 20220190548, 20220204648, 20220204662, 20220207621, 20220207623, 20220207628, 20220207632, 20220207636, 20220207643, 20220204643, 20220204647, 20220204650, 20220204652, 20220204653, 20220204655, 20220204658, 20220204659, 20220204661, 20220204663, 20220204664, 20220207616, 20220207617, 20220207618, 20220207619, 20220207620, 20220207627, 20220207629, 20220207633, 20220207639, 20220261654; UDI-DI 08592654356685. 1ES211-1 - Serial numbers 20220210341, 20220190528, 20220190539, 20220190540, 20220190544, 20220190530, 20220190537, 20220190543, 20220207647, 20220207648, 20220207649, 20220207650, 20220207651, 20220207653, 20220207655, 20220207659, 20220207661, 20220207662, 20220207663, 20220207664, 20220207666, 20220207667, 20220207668, 20220207670, 20220210334, 20220212559, 20220212547, 20220212548, 20220212553, 20220210306, 20220210308, 20220210315, 20220210317, 20220210318, 20220210320, 20220210330, 20220210336, 20220210337, 20220212545, 20220212558; UDI-DI 08592654356678.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FPO
- Firm FEI
- 3008058231
- Firm city
- Zelevcice
- Firm state
- not on file
- Firm country
- Czech Republic
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28