Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: MSK EXTREME MR SCANNER, MODEL AA5000 The intended for use as a diagnostic imaging device to produce axial, sagittal, coronal and oblique images of the internal structures of the leg (excluding the thigh), knee, ankle, foot, forearm, elbow, wrist and hand.

Recall number
Z-1211-2013
Recalling firm
GE Healthcare, LLC
Product
MSK EXTREME MR SCANNER, MODEL AA5000 The intended for use as a diagnostic imaging device to produce axial, sagittal, coronal and oblique images of the internal structures of the leg (excluding the thigh), knee, ankle, foot, forearm, elbow, wrist and hand.
Status
Terminated
Initiated
2013-03-28
Posted
2013-05-01
Terminated
2013-10-25
Reason
Coolant supply lines may leak liquid onto the floor resulting in a slippery surface and fall hazard.
Root cause
Process control
Action
GE Healthcare sent an Urgent Medical Device Correction letter dated April 23, 2013, to all affected customers. The letter informed customers of a potential safety issue due to leakage from gradient coolant lines. Consignees were instructed to monitor the floor for leakage and continue to use if leakage is not observed. If leakage is observed consignees are instructed to stop using the system and notify their local service representative. GE Healthcare will correct all affected scanners by providing an update at no cost. Customers with questions were instructed to contact their local service representataive. For questions regarding this recall call 262-513-4122.
Quantity
70
Distribution
Worldwide Distribution - USA including AL, CA, CO, FL, ID, IL, MA, NH, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI; Worldwide: Argentina, Australia, Austria, Brazil, Canada, Chile, Denmark, Egypt, Finland, Germany, India, Italy, Japan, Korea, Malaysia, Netherlands, Norway, Peru, Poland, Russia, Singapore, South Africa, Spain, Sweeden, Switzerland, UK, Ukraine, and UAE.
Lot, serial or model codes
01072011-001; 01082010-001; 01182010-001; 01182011-001; 01272010-001; 02032011-001; 02112010-001; 02182011-001; 02242010-001; 2262011-001; 03012010-001; 03072010-001; 03132009-001; 03242009-001; 03292011-001; 03302011-001; 04012010-001; 04082010-001; 04192011-001; 04202011-001; 04302010-001; 05072010-001; 05132011-001; 05152009-001; 05242010-001; 05262010-001; 06012008-001; 06042009-001; 06062009-001; 06082010-001; 06182010-001; 06192009-001; 07082010-001; 07092010-001; 07142009-001; 07152011-001; 07272010-001; 08082009-001; 08122009-001; 08122010-001; 08132009-001; 08152010-001; 09072009-001; 09132010-001; 09142010-001; 09172009-001; 09222009-001; 09222010-001; 10012009-001; 10082009-001; 10082010-001; 10142009-001; 10182010-001; 10222010-001; 10252009-001; 10312010-001; 11162010-001; 11242009-001; 11292010-001; 12052010-001; 12072009-001; 12142009-001; 12152009-001; 12162008-001; 12182008-001; 12222010-001
510(k) numbers
["K080048"]
PMA numbers
not on file
Product code
LNH
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register