Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to treat periarticular or intraarticular fractures of the Distal Femur, Proximal Humerus, Proximal Tibia, and the Distal Tibia. Per the operative technique, " A 4.3 mm Drill (REF 702743) is then used to drill the core hole for the locking screw. Using a fluoroscopy, check the correct depth of the drill, and measure the length of the screw. The Drill Sleeve should now be removed, and the correct length 5.0 mm Locking Screw is inserted using the Screwdriver T20 and Screw Holding Sleeve."

Recall number
Z-1210-2013
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to treat periarticular or intraarticular fractures of the Distal Femur, Proximal Humerus, Proximal Tibia, and the Distal Tibia. Per the operative technique, " A 4.3 mm Drill (REF 702743) is then used to drill the core hole for the locking screw. Using a fluoroscopy, check the correct depth of the drill, and measure the length of the screw. The Drill Sleeve should now be removed, and the correct length 5.0 mm Locking Screw is inserted using the Screwdriver T20 and Screw Holding Sleeve."
Status
Terminated
Initiated
2013-03-28
Posted
2013-05-01
Terminated
2015-11-17
Reason
Stryker Orthopaedics has received a report indicating that the scale of the AxSOS Calibrated Drill Bit 4.3,,X263mm, flat , manufactured by Stryker Osteosynthesis, is incorrect. The incorrect scale could result in the surgeon selecting a screw that is 10mm longer than intended.
Root cause
Device Design
Action
Stryker Orthopaedics sent a Urgent Medical Device Removal letter dated March 28, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Please inform users of this Medical Device Removal and pass this notice to all those individuals who need to be aware within your organization. 2. Return all affected products available at your location to Stryker Osteosynthesis or Contact Stryker Customer Service. 3. Complete and sign the enclosed Business Reply Form and fax a copy to: 1-865-252-3635 . 4. Keep a copy of the completed and executed Business Reply Form for your records. Report any adverse events or product quality problems to Stryker Orthopaedics: 1-866-0R-ASSIST. (1-866-672-7747). As we strive for products that meet your expectations for quality and reliability, please do not hesitate to contact us, in case you have any further questions at (201) 831-5970..
Quantity
107 units
Distribution
Nationwide Distribution including the states of AL, AZ, IL, KY, LA, NJ, NC, FL, CA, NM, MO, TN, VA, CO, MA, TX, IN and MT.
Lot, serial or model codes
Catalog number 703615
510(k) numbers
not on file
PMA numbers
not on file
Product code
HTW
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register