Wright Medical Technology Inc: Screwdriver for 3.5mm Screws, DARCO system, instrument for use in orthopaedic implant surgery. REF: DC4261, Non-sterile, Rx only, 1 each, Wright Medical Technology, Inc., 5677 Airline, Road, Arlington, TN 38002

Recall number
Z-1208-2009
Recalling firm
Wright Medical Technology Inc
Product
Screwdriver for 3.5mm Screws, DARCO system, instrument for use in orthopaedic implant surgery. REF: DC4261, Non-sterile, Rx only, 1 each, Wright Medical Technology, Inc., 5677 Airline, Road, Arlington, TN 38002
Status
Terminated
Initiated
2009-02-19
Posted
2009-04-28
Terminated
2010-02-03
Reason
The package insert lists 3 sterilization procedures, 2 of which are not validated. Pre-vacuum steam sterilization is validated.
Root cause
Process design
Action
Wright Medical notified their distributors of the recall on 03/11/2009 by letter explaining the problem and requesting return of the product from their inventory and from their consignees. The letter included a "fax back" response page to ensure the receipt of the notice. The hospitals and surgeons were notified of the recall also on 03/11/2009. All letters were delivered by Fed Ex.
Quantity
682 devices
Distribution
Worldwide distribution: USA, Canada, Germany, France, Belgium, the Netherlands, Italy, Finland, Austria, Denmark, Norway, Turkey, South Africa, United Kingdom, and Australia.
Lot, serial or model codes
Lot numbers: 019598209, 038504328, 058486628, 058507733, 058516752, 067422530, 088538652, 098538652, 107439501, 107442247, 117442247, 0712180, S0701050
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1043534
Firm city
Arlington
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register