Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF): The C304-HIS device features a guide wire to access the vein, a valve to reduce blood loss during the implant procedure, a deflectable catheter to introduce a transvenous device, a catheter dilator to facilitate deflectable catheter passage, and a guide catheter slitter to remove the deflectable catheter. The deflectable catheter is designed for placement of transvenous devices in or near the bundle of His. It features a deflecting distal section, controlled by the deflectable catheter handle, and an out-of plane curve on the distal tip. The body of the deflectable catheter is radiopaque for visibility on fluoroscopy.

Recall number
Z-1206-2022
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Product
The C304-HIS device features a guide wire to access the vein, a valve to reduce blood loss during the implant procedure, a deflectable catheter to introduce a transvenous device, a catheter dilator to facilitate deflectable catheter passage, and a guide catheter slitter to remove the deflectable catheter. The deflectable catheter is designed for placement of transvenous devices in or near the bundle of His. It features a deflecting distal section, controlled by the deflectable catheter handle, and an out-of plane curve on the distal tip. The body of the deflectable catheter is radiopaque for visibility on fluoroscopy.
Status
Open, Classified
Initiated
2022-04-20
Posted
not on file
Terminated
not on file
Reason
The firm's internal processes identified that certain lots may have an improperly sealed barrier that could potentially compromise sterility.
Root cause
Under Investigation by firm
Action
The firm sent out a recall notification starting on 04/20/2022 via UPS 2-day delivery. Customers are asked to return all unused affected product, share the notice with all those who need to be aware within their organization or with any organization where the above-listed products have been transferred, and to complete the response form.
Quantity
1385 devices
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Belgium, Denmark, France, Germany, Ireland, Italy, Netherlands, Poland, Spain, United Arab Emirates, United Kingdom.
Lot, serial or model codes
Model Number: C304-HIS; GTIN: 00763000147013; Lot Numbers: 0010948864, 0010953298, 0010958150, 0010961222, 0010971148, 0011017539, 0011020263, 0011022874, 0011025202, 0011028574, 0011031177
510(k) numbers
["K183493"]
PMA numbers
not on file
Product code
DQY
Firm FEI
2182208
Firm city
Mounds View
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register