Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: CentraLink Data Management System Software Versions: v13x,v14x,v15x The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location. Siemens Healthcare Diagnostics has determined that there is a remote possibility CentraLink may download an order to the ADVIA Automation System without specifying the sample type.

Recall number
Z-1204-2017
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
CentraLink Data Management System Software Versions: v13x,v14x,v15x The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location. Siemens Healthcare Diagnostics has determined that there is a remote possibility CentraLink may download an order to the ADVIA Automation System without specifying the sample type.
Status
Terminated
Initiated
2016-11-29
Posted
not on file
Terminated
2018-06-07
Reason
There is a remote possibility CentraLink may download an order to the ADVIA Automation System without specifying the sample type. This can occur when an order is received from the LIS without a sample type, requiring that the sample type be set in CentraLink based on the sample type of the test in the order.
Root cause
Device Design
Action
Siemens Healthcare Diagnostics, sent an Urgent Medical Device Correction Notice dated November 29, 2016, to all affected customers within the USA, and and issued notices titled "Urgent Field Safety Notices" to their international customers. Effectiveness Check Forms were also sent to the affected customers. Siemens recommends the LIS specify the sample type in the work order. The CentraLink system is designed to operate in an environment in which the sample type is specified in the work order that is downloaded from the LIS. A Siemens Technical Support Representative will contact affected customers to review your CentraLink configuration. If your LIS does not include sample type information in the sample orders downloaded to CentraLink and Pre-Tests assigned by sample type are configured on the ADVIA Automation System your representative will update your configuration. Complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. If you have any questions, please contact your Siemens Customer Care Center or your local Siemens Technical Support Representative. For further questions, please call (914) 631-8000.
Quantity
3,893 units
Distribution
Worldwide Distribution - USA (nationwide) to the states of : AL, AR, AZ, CA, CO, CT FL, GA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, WA and WV; and to the country of Canada and others.
Lot, serial or model codes
Product codes: 10485061, 10484212, 10485064, 10483439, 10471393, 10484552, 10814296, 10815474, 11220480, 10481936, 10481937, 10481938, 10482170, 10482936, 10483437, 10483438, 10483440, 10484211, 10484213, 10485060, 10485062, 10485063, 10703007, 10703008, 10711638, 10711639, 10711640, 10711641, 10711643, 10811596, 10811597, 10811598, 10811599, 10811601, 10811741, 10811742, 10811743, 10814298, 10814877, 10814880, 10814881, 10815694, 10816519, 10816520, 10816521, 10816522, 10816523, 10816524, 10817180, 10817181, 10819760, 10819943, 11219840, 11219841.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register