Thoratec Corp.: Thoratec HeartMate 3 Coring Tool, Sterile EO, Rx Only, REF 10005872, an optional accessory of Thoratec HeartMate 3 Left Ventricular Assist System - Product Usage: The HeartMate 3 Coring Tool is intended for use with the HeartMate 3 Left Ventricular Assist System (LVAS). The Coring Tool provides a means to resect a plug of myocardium from the left ventricle, which allows for HeartMate 3 inflow cannula insertion.

Recall number
Z-1203-2019
Recalling firm
Thoratec Corp.
Product
Thoratec HeartMate 3 Coring Tool, Sterile EO, Rx Only, REF 10005872, an optional accessory of Thoratec HeartMate 3 Left Ventricular Assist System - Product Usage: The HeartMate 3 Coring Tool is intended for use with the HeartMate 3 Left Ventricular Assist System (LVAS). The Coring Tool provides a means to resect a plug of myocardium from the left ventricle, which allows for HeartMate 3 inflow cannula insertion.
Status
Terminated
Initiated
2019-03-01
Posted
2019-04-19
Terminated
2022-08-10
Reason
Small black plastic particle was observed in the left ventricle during implant procedure.
Root cause
Device Design
Action
On 03/01/19, sales representatives started visiting and telephoning consignees informing them to quarantine affected devices. The firm requested that consignees complete and return the Device Retrieval Template. On 03/08/19, Urgent Medical Device Recall notices were hand delivered to all impacted accounts. Customers were advised to stop use, quarantine, and return any units on hand. For patients, who's implant procedure included the use of affected devices, firm recommends close vigilance for potential harm, which could include concerns such as infection or thromboembolism. The coring knife is not impacted. Firm requests that customers complete and return the acknowledgement form. Firm recommends that customer with additional questions call 1-800-456-1477.
Quantity
650
Distribution
US Nationwide distribution in the states of MN, WI, TX, NC, GA, OK, UT, NY, SC, IL, LA, MI, IN, CA, OH, WA. Foreign (OUS): AUSTRIA, BELGIUM, CANADA, ESTONIA, FRANCE, GERMANY, GREECE, INDIA, ITALY, KUWAIT, LIECHTENSTEIN, LITHUANIA, NETHERLANDS, POLAND, SAUDI ARABIA, SINGAPORE, SPAIN, SWITZERLAND, TURKEY, UNITED KINGDOM
Lot, serial or model codes
Lot Numbers: 6675594  6675724  6675730  6677921  6696569  6696580  6696582  6696597  6696601  6696604  6703553  6703561  6703563  6703567  6721603  6721660  6721661  6721663  6721671  6721672  6721673  193496  193497  193499  193501  6425740  6425773  6449684  6675597  6449680  6425751  6425758  6425763  6425768  6449657  6449662  6512124  6677922  6425756  6425766  6425770  6449675  6449681  6449691  6675592  6677919  6677926  6694918  6696574  193500  193502  6425745  6425776  6449663  6449664  6449679  6449682  6449687  6449689  6677924  6677927  6692906  6693957  6696589  193495  193498  6449692  6675591  6675596  6692904  6696599  6512126  6675732  6675734  6675736  6692897  6449677  6692877
510(k) numbers
not on file
PMA numbers
["P160054S005"]
Product code
DSQ
Firm FEI
2916596
Firm city
Pleasanton
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register