Thermo Fisher Scientific: Cascadion SM Clinical Analyzer, REF 99990000 SN Type: 9910

Recall number
Z-1202-2021
Recalling firm
Thermo Fisher Scientific
Product
Cascadion SM Clinical Analyzer, REF 99990000 SN Type: 9910
Status
Terminated
Initiated
2021-02-01
Posted
not on file
Terminated
2024-01-10
Reason
Due to software defect, under certain assay parameters, false Vitamin D results may be reported. The system reports a false result by not quantitating the correct analyte peak in sample chromatogram. The resulting sample would show a Vitamin D-concentration that is an unusual situation for human serum and plasma samples and would indicate a severe Vitamin D deficiency.
Root cause
Software design
Action
On 02/01/2021, the firm sent an "URGENT - MEDICAL DEVICE RECALL" Notification Letter to inform customers that the Recalling Firm has identified that there is a software anomaly in the Analyzer Software subcomponent called Suhaili Service (version 2.0.3 and earlier) that is affecting the Analyzer. Suhaili Service controls raw data processing based on assay configuration data from Sample Prep software. Processed results are sent to Sample Prep software for patient result reporting. This anomaly, as a result of this interface, has a potential risk of reporting wrong patient results for Vitamin D. Customers are informed, As an immediate correction the following instructions must be followed with Cascadion SM Clinical Analyzer Software versions 2.0.2 and earlier until the installation of the new software version 2.1. 1. Please distribute this information immediately to any staff that may be impacted by this issue. 2. Please change the Quick Connect Cartridge Cs on the instrument to new ones. 3. After the cartridge change, if a sample is reported to have concentration of total 25-hydroxy Vitamin D below assay measuring range, follow these instructions to prevent the error from being reported: a. Repeat the sample analysis with a Cascadion SM Clinical Analyzer b. In case the repeated result is below assay measuring range, the sample should be rerun on another assay and platform. 4. Retain a copy of this letter for your laboratory records. 5. As appropriate, contact your Medical Professional for evaluation of further action. 6. Please, fill out the MEDICAL DEVICE RECALL- Response Form and return it within 5 days of the date of this letter to your manufacturer as instructed in the form and as listed: Email: vigilance.clinical.fi@thermofisher.com The actions to be taken by the Recalling Firm: 1. The Recalling Firm has informed the appropriate Regulatory Agencies. 2. When, available, the final corrective action will be to update Clinical Analyzer software v2.1
Quantity
19 systems (1 system in U.S. and 10 system O.U.S.)
Distribution
U.S.: Texas O.U.S.: Belgium, France, Germany, Italy, Spain, Switzerland and United Kingdom
Lot, serial or model codes
Model Number 99990000 GTIN: 06438153004925 Serial Numbers: CA100077, CA000055, CA100073, CA100074, CA000051, CA100062, CA100065, CA100069, CA100070 CA100063 CA000042, CA100071 CA100060, CA100059, CA000058, CA000053, CA000056, CA100061, CA100066
510(k) numbers
not on file
PMA numbers
not on file
Product code
DOP
Firm FEI
3007262384
Firm city
Vantaa
Firm state
not on file
Firm country
Finland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register