Varian Medical Systems, Inc.: Nucletron Adapter and Transfer Guide Tube Set, part number AL 13005001, set of 3 adapters, components used with the VariSource" IX HDR Afterloader. Product Usage: The adapter is part of the VariSource Afterloader System, a computer controlled remote HDR Afterloader used to place a high activity radioactive source within a needle(s) or applicator(s) which have previously been placed for a specific clinical purpose in a patient.

Recall number
Z-1200-2012
Recalling firm
Varian Medical Systems, Inc.
Product
Nucletron Adapter and Transfer Guide Tube Set, part number AL 13005001, set of 3 adapters, components used with the VariSource" IX HDR Afterloader. Product Usage: The adapter is part of the VariSource Afterloader System, a computer controlled remote HDR Afterloader used to place a high activity radioactive source within a needle(s) or applicator(s) which have previously been placed for a specific clinical purpose in a patient.
Status
Terminated
Initiated
2012-01-05
Posted
2012-03-12
Terminated
2012-11-21
Reason
Components of medical device used in brachytherapy may be defective and cause a shifted dose distribution to patients.
Root cause
Device Design
Action
Varian Medical Systems sent an Urgent Medical Device Correction - Urgent Field Safety Notice letter dated January 5, 2012 to all affected consignees. The letter identified the affected products, description of problem and recommended user actions to be taken. Consignees were requested to cease use of the affected products and contact Varian BrachyTherapy Helpdesk immediately to arrange return of their affected parts. Consignees were requested to return the enclosed proof of notification form. Consignees were instructed to advise the appropriate personnel working in their radiotherapy department of the content of this letter. For future reference, this document will be posted to the Varian Customer Support website: http://www.MyVarian.com. For further clarification contact your local Varian District or Regional Manager.
Quantity
122 total units
Distribution
Worldwide Distribution - USA (nationwide) Including the states of CA, FL, IN, KY, MI, MN, MO, OH, TX, VA and the countries of Canada, Japan and Sweden.
Lot, serial or model codes
All lots of the components are recalled. Components are used as adapters for the Varian Afterloaders with the following serial numbers:  H600097 H600254 H600255 H600268 H600274 H600341 H600345 H600352 H600377 H600381 H600416 H600435 H600437 H600462 H600513 H600524 H600538 H600546 H600548 H600583
510(k) numbers
["K071467"]
PMA numbers
not on file
Product code
JAQ
Firm FEI
3000206172
Firm city
Charlottesville
Firm state
VA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register