FDA Medical Device Recalls (openFDA)
Acumed LLC: 6-hole Pubic Symphysis plate (PN 70-0451). The pelvic bone plates for the pelvic ring are a series of plates with varying lengths that function as internal fixation devices for fractures, fusions and osteotomies of the pelvic ring. The plates are strategically pre-contoured where beneficial to the application, and they are secured to the bone with 3.5mm non-locking screws. The Acumed pelvic ring plate is the Pubic Symphysis Plate. The Acumed Pelvic Plating System is intended for use by surgeons with orthopedic training and knowledge of the indications and techniques required for fixation. The device is to be implanted by the surgeon in a sterile operating room setting.
- Recall number
- Z-1199-2017
- Recalling firm
- Acumed LLC
- Product
- 6-hole Pubic Symphysis plate (PN 70-0451). The pelvic bone plates for the pelvic ring are a series of plates with varying lengths that function as internal fixation devices for fractures, fusions and osteotomies of the pelvic ring. The plates are strategically pre-contoured where beneficial to the application, and they are secured to the bone with 3.5mm non-locking screws. The Acumed pelvic ring plate is the Pubic Symphysis Plate. The Acumed Pelvic Plating System is intended for use by surgeons with orthopedic training and knowledge of the indications and techniques required for fixation. The device is to be implanted by the surgeon in a sterile operating room setting.
- Status
- Terminated
- Initiated
- 2017-01-19
- Posted
- 2017-02-14
- Terminated
- 2017-05-08
- Reason
- The 6-hole Pubic Symphysis plate (PN 70-0451) is specified to be made of Commercially Pure Titanium (ASTM F67 - CP Grade 4). However, batch 383813 was manufactured out of Titanium Alloy (Ti-6Al- 4V ELI per ASTM F135 Ti alloy) and was distributed.
- Root cause
- Mixed-up of materials/components
- Action
- Acumed sent an Urgent Medical Device Recall letter dated January 16, 2017, to all affected customers on January 19, 2017. Consignees were instructed to follow special instructions and send survey to RecallNotification@acumed.net. Domestic customers with questions or need to return products to Acumed, please contact Acumed Agent Inventory via email AgentInventory@acumed.net to obtain a Return Material Authorization (RMA) number or contact your agent directly. International customers with questions or need to return products to Acumed, please contact Acumed Business Service via email at BusinessServices@acumed.net or via phone 877-627-9957, choose Option 4 (US and Canada), or call 001-503-627-9957, choose option 4 (outside of US and Canada) have to obtain a Return Material Authorization (RMA) number.
- Quantity
- 15 units
- Distribution
- Worldwide Distribution - Costa Rica, Malaysia, Mexico, South Africa, and Spain
- Lot, serial or model codes
- Lot number: 383813
- 510(k) numbers
- ["K122538"]
- PMA numbers
- not on file
- Product code
- HRS
- Firm FEI
- 1000125930
- Firm city
- Hillsboro
- Firm state
- OR
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28