Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, LP: Medline medical procedure convenience kits labeled as: 1) DEPAUL BASIC NEURO-LF, Kit SKU DYNJ20028L; 2) NEURO MINOR PACK, Kit SKU DYNJ46546M; 3) NEURO SHUNT PACK, Kit SKU DYNJ54396K; 4) NEURO PACK, Kit SKU DYNJ59397Q; 5) NEURO BASIC, Kit SKU DYNJ904052L; 6) PK NEURO SHUNT BUMC, Kit SKU DYNJT6428.

Recall number
Z-1196-2026
Recalling firm
Medline Industries, LP
Product
Medline medical procedure convenience kits labeled as: 1) DEPAUL BASIC NEURO-LF, Kit SKU DYNJ20028L; 2) NEURO MINOR PACK, Kit SKU DYNJ46546M; 3) NEURO SHUNT PACK, Kit SKU DYNJ54396K; 4) NEURO PACK, Kit SKU DYNJ59397Q; 5) NEURO BASIC, Kit SKU DYNJ904052L; 6) PK NEURO SHUNT BUMC, Kit SKU DYNJT6428.
Status
Open, Classified
Initiated
2025-12-12
Posted
2026-01-28
Terminated
not on file
Reason
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root cause
Nonconforming Material/Component
Action
Medline Industries issued a URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/12/2025 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-25-251-FGX2 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity
236 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
Medline Kit SKU DYNJ20028L: UDI/DI 10198459228995 (EA) 40198459228996 (CS), Lot Number 25LMA824; Medline Kit SKU DYNJ46546M: UDI/DI 10198459526404 (EA) 40198459526405 (CS), Lot Number 25KMH940; Medline Kit SKU DYNJ54396K: UDI/DI 10198459443138 (EA) 40198459443139 (CS), Lot Number 25KMD804; Medline Kit SKU DYNJ59397Q: UDI/DI 10198459495427 (EA) 40198459495428 (CS), Lot Number 25KMC287; Medline Kit SKU DYNJ904052L: UDI/DI 10198459231339 (EA) 40198459231330 (CS), Lot Number 25JDB033; Medline Kit SKU DYNJT6428: UDI/DI 10198459553660 (EA) 40198459553661 (CS), Lot Number 25KMD463.
510(k) numbers
not on file
PMA numbers
not on file
Product code
OJG
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register