Skip to the content

FDA Medical Device Recalls (openFDA)

Argon Medical Devices, Inc: T-LOK BONE MARROW BIOPSY NEEDLE 8G x 4 EXT CAN; T-LOK BONE MARROW BIOPSY NEEDLE 11G x 4 EXT CAN Intended for the purpose of harvesting bone and/or bone marrow specimens

Recall number
Z-1195-2017
Recalling firm
Argon Medical Devices, Inc
Product
T-LOK BONE MARROW BIOPSY NEEDLE 8G x 4 EXT CAN; T-LOK BONE MARROW BIOPSY NEEDLE 11G x 4 EXT CAN Intended for the purpose of harvesting bone and/or bone marrow specimens
Status
Terminated
Initiated
2017-01-18
Posted
2017-02-13
Terminated
2017-05-17
Reason
The recalling firm received a complaint from one of their customers of a potential defect in the forming of the primary package of the blister that may compromise the sterility of the product inside the tray.
Root cause
Equipment maintenance
Action
Argon Medical Devices, Inc. is sending an Urgent Product Recall Notice letter dated January 17, 2017, by registered mail to all affected customers. Customers were asked to determine if their inventory shows the package defect and discontinue use of the device and promptly return the affected product. For questions regarding this recall call 903-403-0749.
Quantity
772 boxes
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Switzerland, Canada, Australia, New Zealand, Ecuador, Colombia, and Japan.
Lot, serial or model codes
lots: 11157523, 11158179, 11158180, 11159101, 11159729
510(k) numbers
["K980196"]
PMA numbers
not on file
Product code
KNW
Firm FEI
1625425
Firm city
Athens
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register