Skip to the content

FDA Medical Device Recalls (openFDA)

Arthrex, Inc.: Arthrex IBalance TKA Tibial Tray Size 2 - Catalog # AR-503-TTTC, Size 3 - Catalog # AR-503-TTTD, Size 4 -Catalog # AR-503-TTTE, Size 5 - Catalog # AR-503-TTTF, Size 6 - Catalog # AR-503-TTTG , Size 7 -Catalog # AR-503-TTTH, and Size 8 - Catalog # AR-503-TTTJ. Intended for implantation with bone cement, with the exception of the porous coated femoral components which can be used cemented or uncemented (biological fixation).

Recall number
Z-1195-2016
Recalling firm
Arthrex, Inc.
Product
Arthrex IBalance TKA Tibial Tray Size 2 - Catalog # AR-503-TTTC, Size 3 - Catalog # AR-503-TTTD, Size 4 -Catalog # AR-503-TTTE, Size 5 - Catalog # AR-503-TTTF, Size 6 - Catalog # AR-503-TTTG , Size 7 -Catalog # AR-503-TTTH, and Size 8 - Catalog # AR-503-TTTJ. Intended for implantation with bone cement, with the exception of the porous coated femoral components which can be used cemented or uncemented (biological fixation).
Status
Terminated
Initiated
2015-12-17
Posted
not on file
Terminated
2016-12-14
Reason
Devices reported to have a smooth texture to the outer surface of the metal, which differed from a previous that had a textured outer surface.
Root cause
Nonconforming Material/Component
Action
On February 5, 2016 Arthrex sent Urgent Medical Device Voluntary Recall letters to their distributors and Surgeons/Medical Facilities reporting the removal. The letter identified the reason for the recall, the potential patient impact, and actions to be taken. Users were to discontinue use of the identified affected devices and follow the instructions provided to return the affected devices that have not been implanted. To return the devices, contact Arthrex at 1-866-267-9138 or by fax at 1-239-591-6943 or via e-mail at Complaints@Arthrex.com. Customers must complete and return the Acknowledgement and Receipt.
Quantity
2,378 units.
Distribution
Nationwide Distribution -- AR, AZ, CA, CO, DC, DE, FL, GA, IL, IN, KS, MA, MD, MI, MN, MO, NC, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, WA, WI and District of Columbia.
Lot, serial or model codes
Size 2 Unique Device Identifier: 00888867106703 Batch #'s: 927692, 915207, 982771, 979834, 982774, 1206164, 1206165, 1206167, 1206166, 1227388, 1227387, 1257195, 1233620, 1289648, 1233621, 1289653, 1289736, 1289654, 1336430, 1336431, 1305721, 1306332, 1352062, 1336433, 1336436, 1352058, 1352063, 1376104, 1376106, and 1376107.  Size 3 Unique Device Identifier: 00888867106710 Batch #'s: 780838, 952789, 1206169, 1206170, 1206168, 1191337, 1191337, 1272728, 1297048, 1297044, 1297049, 1272730, 1272729, 1307262, 1297060, 1297057, 1297365, 1307263, 1336445, 1307267, 1336451, 1336448, 1336451, 1307265, 1336453, 1346958, 1346962, 1346982, 1346994, 1346965, 1336452, 1379085, 1379104, 1379110, 1379113, 1379114, 1346999, and 1379084.  Size 4 Unique Device Identifier: 00888867106727 Batch #'s: 894339, 913623, 874733, 876482, 876478, 1191338, 1206172, 1206173, 1297075, 1297084, 1315459, 1297086, 1297366, 13115453, 1315454, 1315459, 1315462, 1315464, 1336466, 1336465, 1336467, 1336455, 1336468, 13336469, 1356527, 1356530, 1356531, 1356529, 1356574, 1379205, 1379203, 1379204, 1379206, 1379200, 1356576, 1379196, and 1379201. Size 5 Unique Device Identifier: 00888867106734 Batch #'s:780840, 971202, 1206176, 1206177, 1356586, 1322523, 1380216, 1380223, 1380224, 1356588, 1380221, 1380226, 1555653, and 1555654. Size 6 Unique Device Identifier: 00888867106741 Batch #'s:780841, 970368, 1206178, 1206179, 1322459, 1322460, 1322524, 1357785, 1380243, 1380244, 1380245, 1380246, 1357786, and 1380242. Size 7 Unique Device Identifier: 00888867106758 Batch #'s: 780842, 982767, 984546, 1206180, 1337230, 1206181, 1337234, 1337237, 1337240, 1337241, 1337256, 1358297, 1358299, 1358298, 1358300, 1358302, 1380708, 1380709, and 1380710. Size 8 Unique Device Identifier: 00888867106765 Batch #'s: 913624, 970865, 957056, 1191342, 1224471, 120800, 1191341, 1191339, 876484, 876485, 1228632, 1251021, 1292154, 1251022, 1281699, 1292161, 1293756, 1293758, 1307294, 1307295, 1308538, 1337262, 1337300, 1337303, 1347012, 1347013, 1337325, 1347014, 1380743, 1347025, 1380747, 1380748, and 1380742.
510(k) numbers
["K081127"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1220246
Firm city
Naples
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register