MRL, Inc., A Welch Allyn Company: Welch Allyn PIC 50 Portable Intensive Care System Multi-Parameter Monitor/Defibrillator; a portable, 12V internal battery powered defibrillator; Manufactured by MRL Inc., A Welch Allyn Company, 1000 Asbury Drive, Buffalo Grove, Illinois 60089

Recall number
Z-1195-06
Recalling firm
MRL, Inc., A Welch Allyn Company
Product
Welch Allyn PIC 50 Portable Intensive Care System Multi-Parameter Monitor/Defibrillator; a portable, 12V internal battery powered defibrillator; Manufactured by MRL Inc., A Welch Allyn Company, 1000 Asbury Drive, Buffalo Grove, Illinois 60089
Status
Terminated
Initiated
2006-06-30
Posted
2006-07-08
Terminated
2007-11-20
Reason
The Welch Allyn PIC 50 Defibrillators may display a 'Defib Comm' or 'Pace Comm' error message on the device display during use which may result in a terminal failure of the device to analyze the patient's ECG and deliver the appropriate therapy.
Root cause
Other
Action
MRL sent Urgent Medical Device Recall letters dated 6/30/06 via certified mail to the direct accounts informing them of an intermittent connection between a pin and socket on a circuit board inside of the device that could result in a malfunction that prevents or unacceptably delays the delivery of a defibrillating shock, which possibly could result in failure to resuscitate the patient. The accounts were provided a list of the affected units' parts/serial numbers and requested to inspect their PIC 50 units for the affected serial numbers and return the identified units for servicing. The accounts were also instructed to respond to the notification within 5 working days if the unit has displayed the Defib Comm error, and within 30 days if it has not displayed the Defib Comm error by fax or mail. A loaner PIC 50 will then be sent to the consignee with instructions on how to return their unit for correction.
Quantity
673 units - USA, 511 units - International
Distribution
Worldwide-including USA, Puerto Rico, Australia, Belgium, Canada, China, Cyprus, France, Germany, India, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Malaysia, Poland, Portugal, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Sultanate of Oman, Switzerland, United Arab Emirates, United Kingdom and Vietnam.
Lot, serial or model codes
Part numbers 971081, 971082, 971083 and 971084; serial numbers 80104, 80119, 80120, 80204, 80211, 80212, 80429, 80432, 80569, 80627, 80661, 80685, 80880, 80964, 81056, 81134, 81263, 81293/81295, 81303, 81445/ 81448, 81733, 81738, 81739, 81744/81752, 81757, 81761/81781, 81802/81807, 81824, 81825, 81835, 81845, 81846, 81852, 81864, 81865, 81868, 81869, 81884/81886, 81892/81897, 81918/81955, 81957, 81983/81990, 81996/82025, 82027/82033, 82035/82039, 82043, 82044, 82061/82080, 82084, 82097/82111, 82122/82128, 82133, 82136/82143, 82150/82169, 82172/82204, 82245, 82252/82273, 82283/82286, 82307/82330, 82340/82349, 82405/82426, 82430, 82460/82490, 82498/82553, 82555/82565, 82568/82570, 82612/82631, 82654/82673, 82689, 82690, 82773/82813, 82827/82830, 82852, 82856/82859, 82906, 82936, 82937, 82958/82977, 82983/83013, 83060/83071, 83083/83090, 83093, 83142/83147, 83153/83172, 83198/83239, 83255, 83368/83370, 83400/83407, 83414/83418, 83420/83429, 83440/83443, 83499/83517, 83578/83580, 83633, 83644, 83645, 83654, 85001/85005, 85009/85013, 85026/85029, 85032/85052, 85073/85079, 85093, 85094, 85110, 85111, 85143/85157, 85187/85201, 85223, 85224, 85226/85238, 85244/85249, 85329, 85332/85334, 85337/85350, 85364, 85365, 85368/85370, 85381, 85388/85398, 85402/85404, 85410/85414, 85420/85433, 85441, 85505/85515, 85517/85520, 85522/85530, 85534, 85552, 85554/85557, 85587/85589, 85602/85607, 85619/85623, 85630, 85645/85655, 85658/85667, 85686, 85728/85734, 85755/85757, 85761/85766, 85771/85785, 85787/85793, 85810, 85812/85823, 85825/85838, 85856/85872, 85877/85884, 85888/85902, 85931/85933, 85945, 85955, 85964, 85967/85970, 86017, 86018, 86024/86029, 86055/86057, 86103, 86104, 86118, 86141/86147, 86149, 86150, 86222/86228, 86230, 86245, 86256/86258, 86261/86268, 86281/86283, 86289/86293, 86295, 86296, 86299, 86305, 86306, 86314, 86315, 86320/86322, 86324/86328, 86334, 86335, 86358, 86359, 86385, 86387, 86389/86399, 86400, 86411, 86412, 86419, 86420, 86439, 86440
510(k) numbers
["K974034"]
PMA numbers
not on file
Product code
MJK
Firm FEI
1418729
Firm city
Buffalo Grove
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register