Roche Diagnostics Corporation: Albumin BCP Product Usage: ALBP is an in vitro test used for the quantitative determination of albumin in human serum and plasma on Roche/Hitachi cobas c systems.

Recall number
Z-1194-2019
Recalling firm
Roche Diagnostics Corporation
Product
Albumin BCP Product Usage: ALBP is an in vitro test used for the quantitative determination of albumin in human serum and plasma on Roche/Hitachi cobas c systems.
Status
Terminated
Initiated
2019-03-18
Posted
not on file
Terminated
2020-08-28
Reason
The reagent lot is exhibiting calibration failures, QC imprecision, and elevated QC recovery due to elevated pH level causing higher absorbance values.
Root cause
Process control
Action
Urgent Medical Device Correction notification letters dated 3/18/19 were sent to customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to: Stop using and discard all Albumin BCP reagent cassettes from lot number 35651401 (catalog number 05975573190) remaining in your inventory according to your local guidelines. Actions Required " Complete all sections of the enclosed fax form and fax it to 1-833-254-2600 or email it to Roche10060@stericycle.com even if you are not requesting product replacement. " Discard remaining affected product in your inventory according to local guidelines. " File this Urgent Medical Device Correction (UMDC) for future reference.
Quantity
774
Distribution
US Nationwide distribution in the states of: GA, IL, KY, MI, and TX.
Lot, serial or model codes
Catalog Number 05975573190 Lot Number 35651401
510(k) numbers
["K844426"]
PMA numbers
not on file
Product code
CJW
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register