Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Spine, Inc.: Universal Clamp System (IFU reference: SNA027-N-90001). Provides the stability required for solid bone fusion and helps reduce spinal deformities and fractures.

Recall number
Z-1189-2016
Recalling firm
Zimmer Spine, Inc.
Product
Universal Clamp System (IFU reference: SNA027-N-90001). Provides the stability required for solid bone fusion and helps reduce spinal deformities and fractures.
Status
Terminated
Initiated
2016-02-12
Posted
not on file
Terminated
2016-08-11
Reason
Zimmer Biomet is initiating a Medical Device Correction Notice to provide updated cleaning and disinfecting instructions for use of the Instinct Java System (IFU reference: 046WAN0000T) and Universal Clamp System (IFU reference: SNA027-N-90001).
Root cause
Process control
Action
Consignees were sent on 2-12-2016 a Zimmer Biomet "Urgent Medical Device Correction Notice" letter dated February 12, 2016. The letter was either addressed to Hospital Representative or to Distributors, Sales Representatives, and Distribution Operations Managers. The letters described the problem and the product involved in the recall. Advised consignees to immediately utilize the updated and attached Instructions For Use. Requested consignees to complete the "Acknowledgement of Field Notification" form. For questions contact Customer Service at 1-866-774-6368 or e-mail corporatequality.postmarket@zimmerbiomet.com.
Quantity
3484
Distribution
US (nationwide)-including the states of AL, AZ, AR, CA, CO, FL, GA, ID, IL, IN, IA, LA, ,MD, MA, MI, MN,MO, NV, NH, NY, NC, OH, OK,. PA, TN,TX, WA, and WI.
Lot, serial or model codes
All lots
510(k) numbers
not on file
PMA numbers
not on file
Product code
NKB
Firm FEI
1000115775
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register