Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Navigation, Inc.: Left Long Tactile Probe, Part #962009S, Medtronic Navigation, Louisville, CO 80027. Trackable hand held instruments used in spine surgeries for preparation of the pedicle. These instruments are used in conjunction with the Medtronic Navigation StealthStation platform and various software applications.

Recall number
Z-1189-2012
Recalling firm
Medtronic Navigation, Inc.
Product
Left Long Tactile Probe, Part #962009S, Medtronic Navigation, Louisville, CO 80027. Trackable hand held instruments used in spine surgeries for preparation of the pedicle. These instruments are used in conjunction with the Medtronic Navigation StealthStation platform and various software applications.
Status
Terminated
Initiated
2012-02-06
Posted
2012-03-08
Terminated
2012-03-26
Reason
When subjected to bending, twisting, and malleting, the probe tips break off when used in the sacrum for which they are not indicated for use. These probes are approved for use in the lumbar vertebra.
Root cause
Device Design
Action
An Urgent - Product Safety Notice customer letter was sent via FED EX on 2/6/12 and 2/6/12 to direct consignees in the US identifying the affected product and the issue with it. The letter states that Medtronic Navigation is removing the product from clinical use and will provide a sales credit to customers. In the letters to Medtronic field personnel, it states that local sales and service personnel will be identified to remove the affected product from customer sites and return them for scrapping. Customers are to contact a sales representative if they have any questions regarding the letter.
Quantity
6 units
Distribution
Worldwide Distribution -- US and Republic of South Korea.
Lot, serial or model codes
Part Number 962009S: Lot Numbers 090106, 090407, 071219
510(k) numbers
["K990214"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1000517638
Firm city
Louisville
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register