Skip to the content

FDA Medical Device Recalls (openFDA)

Maquet Medical Systems USA: BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code 70106.9077, BEQ-HLS 7050 USA; HLS Set Advanced 7.0, Product Code 70106.9078. Part of the CARDIOHELP System, a pre-connected set used for extracorporeal respiratory and/or cardiovascular support.

Recall number
Z-1188-2023
Recalling firm
Maquet Medical Systems USA
Product
BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code 70106.9077, BEQ-HLS 7050 USA; HLS Set Advanced 7.0, Product Code 70106.9078. Part of the CARDIOHELP System, a pre-connected set used for extracorporeal respiratory and/or cardiovascular support.
Status
Open, Classified
Initiated
2023-02-13
Posted
2023-02-28
Terminated
not on file
Reason
Firm has initiated a removal of the product due to insufficient evidence of packaging sterility. Exposure to a non-sterile medical device may result in inflammation, infection, sepsis, or ischemia.
Root cause
Process control
Action
On February 13, 2023, the firm notified customers via Urgent Medical Device Correction letters. Customers were informed of the potential packaging issues and the associated risks to health. Customers were instructed to stop use and return all unused and unexpired affected product for credit. If customers have no alternative product, they may choose to continue use of the product, but they must perform a visual inspection of the primary packaging, checking for visible stress marks or damage to the packaging. If there are visible stress marks in the packaging, the customer should not use the product. Please note that there is a possibility that packaging defects may not be detectable by the naked eye. If you have any questions, please contact your Getinge representative or call the Getinge Customer Support at (888) 9GETUSA / (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Quantity
10,587 units (US); 94,987 (worldwide)
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czechia, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Mauritius, Myanmar, Mexico, Moldova, Morocco, Namibia, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic of North Macedonia, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, The Republic of Korea, The Russian Federation, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Vietnam.
Lot, serial or model codes
HLS Set Advanced 5.0, Product Code 70106.9077 UDI-DI 04058863076355 HLS Set Advanced 7.0, Product Code 70106.9078 UDI-DI 04058863080383 All lots/batches affected
510(k) numbers
["K112360"]
PMA numbers
not on file
Product code
DTZ
Firm FEI
3008355164
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register