Skip to the content

FDA Medical Device Recalls (openFDA)

Encore Medical, LP: EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6, 22MM, REF: 346-22-706; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 7, 22MM, REF: 346-22-707; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 8, 22MM, REF: 346-22-708; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 9, 22MM, REF: 346-22-709; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 10, 22MM, REF: 346-22-710; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 11, 22MM, REF: 346-22-711; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 5 BRIDGE UP, 22MM, REF: 346-22-755; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6 BRIDGE DOWN, 22MM, REF: 346-22-766; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 2, 25MM, REF: 346-25-702; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 3, 25MM, REF: 346-25-703; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 4, 25MM, REF: 346-25-704; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 5, 25MM, REF: 346-25-705; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 5 BRIDGE UP, 25MM, REF: 346-25-755; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6 BRIDGE DOWN, 25MM, REF: 346-25-766

Recall number
Z-1187-2023
Recalling firm
Encore Medical, LP
Product
EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6, 22MM, REF: 346-22-706; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 7, 22MM, REF: 346-22-707; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 8, 22MM, REF: 346-22-708; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 9, 22MM, REF: 346-22-709; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 10, 22MM, REF: 346-22-710; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 11, 22MM, REF: 346-22-711; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 5 BRIDGE UP, 22MM, REF: 346-22-755; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6 BRIDGE DOWN, 22MM, REF: 346-22-766; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 2, 25MM, REF: 346-25-702; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 3, 25MM, REF: 346-25-703; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 4, 25MM, REF: 346-25-704; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 5, 25MM, REF: 346-25-705; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 5 BRIDGE UP, 25MM, REF: 346-25-755; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6 BRIDGE DOWN, 25MM, REF: 346-25-766
Status
Open, Classified
Initiated
2022-12-22
Posted
2023-02-27
Terminated
not on file
Reason
Tibial insert implant large boxes were labeled small and the small boxes were labeled large. If the incorrect labeling is identified this could lead to surgical delay and if the surgeon does not recognize the size discrepancy between the insert label and the pin label, the surgeon may implant the incorrect pin size.
Root cause
Labeling Change Control
Action
On 12/22/22 recall notices were emailed to surgical agents who were asked to do the following: 1) Pass on the notice to all those within their organization or to any organization where potentially affected products have been transferred. 2) Return affected products. 3) Complete and return the Consignee Reconciliation notice. The following number was provided: Customer Service, 1-800-456-8696 On 2/8/23 recall notices were emailed with an expanded scope.
Quantity
216
Distribution
US: MN, IL, IN, NY, FL
Lot, serial or model codes
REF/UDI-DI/Lot: 346-22-706/ 00190446219378/484V1013, 484V1014, 484V1015, 484V1016, 484V1017, 484V1018; 346-22-707/ 00190446219385/485V1001, 485V1002, 485V1003, 485V1004; 346-22-708/ 00190446219392/486V1005, 486V1006, 486V1007, 486V1008; 346-22-709/ 00190446219408/487V1001, 487V1002, 487V1003, 487V1004; 346-22-710/ 00190446219415/488V1003, 488V1004, 488V1005, 488V1006; 346-22-711/ 00190446219422/489V1003, 489V1004, 489V1005, 489V1006; 346-22-755/ 00190446219439/490V1001, 490V1002, 490V1003, 490V1004; 346-22-766/ 00190446219477/491V1002, 491V1003, 491V1004; 346-25-702/ 00190446219507/492V1002, 492V1003, 492V1004, 492V1005; 346-25-703/ 00190446219569/493V1001, 493V1002, 493V1003, 493V1004; 346-25-704/ 00190446219583/494V1001, 494V1002, 494V1003, 494V1004; 346-25-705/ 00190446219590/495V1001, 495V1002, 495V1003, 495V1004; 346-25-755/ 00190446219781/ 502V1001, 502V1002, 502V1003; 346-25-766/ 00190446219798/503V1003, 503V1004, 503V1005, 503V1006, 503V1007;
510(k) numbers
["K180930"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1000116912
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register