Skip to the content

FDA Medical Device Recalls (openFDA)

Covidien Medical Products (Shanghai) Manufacturing L.L.C.: Short Percutaneous Antenna with Thermosphere" Technology Standard Percutaneous Antenna with Thermosphere" Technology; The product is packaged in a blister with a tyvek lid, contained in a display box. The numbers of units per carton is 1. Long Percutaneous Antenna with Thermosphere" Technology Product Usage: The Covidien Emprint Ablation System is intended for use in percutaneous,laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of non-resectable liver tumors. The CovidienEmprint Ablation System is not intended for use in cardiac procedures

Recall number
Z-1187-2016
Recalling firm
Covidien Medical Products (Shanghai) Manufacturing L.L.C.
Product
Short Percutaneous Antenna with Thermosphere" Technology Standard Percutaneous Antenna with Thermosphere" Technology; The product is packaged in a blister with a tyvek lid, contained in a display box. The numbers of units per carton is 1. Long Percutaneous Antenna with Thermosphere" Technology Product Usage: The Covidien Emprint Ablation System is intended for use in percutaneous,laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of non-resectable liver tumors. The CovidienEmprint Ablation System is not intended for use in cardiac procedures
Status
Terminated
Initiated
2016-02-23
Posted
2016-03-17
Terminated
2017-06-21
Reason
Medtronic is voluntarily recalling specific item codes and production lots of Covidien Emprint" Percutaneous Antenna with Thermosphere" Technology due to the potential for the generator to shut down in the precence of high temperature measurements in the Emprint" Percutaneous Antenna.
Root cause
Nonconforming Material/Component
Action
Medtronic sent an Urgent Medical Device Recall letter dated February 23, 2016 to all affected customers via Federal Express or Certified Mail. The letter identified the affected product, problem and actions. Customers were advised to quarantine and discontinue use of the affected product and to Return product. For questions or concerns, contact your Medtronic representative or Customer Service at (800) 882-5878.
Quantity
4,920 units
Distribution
Worldwide Distribution - US Nationwide, VA/govt/military consignees.
Lot, serial or model codes
Model Numbers: CA15L1, CA20L1 and CA30L1
510(k) numbers
["K133821"]
PMA numbers
not on file
Product code
NEY
Firm FEI
3006451981
Firm city
not on file
Firm state
not on file
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.