FDA Medical Device Recalls (openFDA)
Brainlab AG: ExacTrac Dynamic (Medical Charged-Particle Radiation Therapy System).
- Recall number
- Z-1185-2021
- Recalling firm
- Brainlab AG
- Product
- ExacTrac Dynamic (Medical Charged-Particle Radiation Therapy System).
- Status
- Terminated
- Initiated
- 2021-03-01
- Posted
- not on file
- Terminated
- 2023-03-20
- Reason
- In case of a failed automatic marker detection, a software error causes parts of the display to incorrectly behave as if the current patient position is within predefined tolerances and may allow the user to proceed to treatment despite potentially exceeding shift values.
- Root cause
- Software design
- Action
- The Field Safety Notice / Product Notification Letter CAPA-20210210-002387 was distributed via email on 03/01/2021. An active reply by the consignees will be followed, if necessary via further different communication media. For future customers/facilities/systems (new installations or updates/upgrades): As of February 23, 2021: - the Field Safety Notice / product notification information is handed over to the customer during installation (before clinical use) by the according Brainlab Service representative, - a confirmation is required that the contained User Corrective Actions will be implemented before clinical use by the facility (to be adhered to by the users).
- Quantity
- 73 ExacTrac Dynamic systems (versions 1.0.0, 1.0.1, 1.0.2) have been manufactured and distributed
- Distribution
- USA: (CA, NJ, AZ, AR, NY, IN, OR, NC), Australia, Austria, Belgium, Denmark, France, Germany, Hong Kong, Italy, Slovakia, Spain, Thailand, Turkey, United Kingdom, United Arab Emirates.
- Lot, serial or model codes
- Model/catalogue numbers: 20910-06 ETD Implanted lic. UDI: The GTIN for ExacTrac Dynamic version 1.0 is 04056481142315. GMDN code: 40887. Software revisions of affected ExacTrac Dynamic software: versions 1.0.0, 1.0.1, 1.0.2.
- 510(k) numbers
- ["K201276"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 3002619595
- Firm city
- Munich
- Firm state
- not on file
- Firm country
- Germany
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28