Skip to the content

FDA Medical Device Recalls (openFDA)

Medical Components, Inc dba MedComp: Power Injectable CT Port insertion kits - Product Usage: is an implantable access device designed to provide repeated access to the vascular system.

Recall number
Z-1184-2021
Recalling firm
Medical Components, Inc dba MedComp
Product
Power Injectable CT Port insertion kits - Product Usage: is an implantable access device designed to provide repeated access to the vascular system.
Status
Open, Classified
Initiated
2021-01-13
Posted
2021-03-08
Terminated
not on file
Reason
The 5F Dignity CT Port kits were packaged with the incorrect size port. The kits include a 6.6F port instead of the correct 5F port. The inability to connect the catheter lumen to the port stem because of the incorrect size may not be noted until well into the procedure, after the catheter lumen is placed in the vessel. An alternate port will need to be obtained prolonging the procedure. A prolonged insertion procedure increases any risk associated with a surgical procedure.
Root cause
Employee error
Action
On 1/13/2021 and 02/10/2021, Medical Components issued an Urgent Medical Device Removal notification to customers via letter notifying customers the 5F Dignity CT Port kits were packaged with the incorrect size port. The kits include a 6.6F port instead of the correct 5F port. (updated 3/11/2021)
Quantity
96 pieces total (updated 3/11/2021 additional 72 pieces)
Distribution
US Nationwide distribution in the states of CA, DC, GA, IL, IN, MA, MI, NJ, NM, NY, OH, PA, TX. Updated 3/11/2021
Lot, serial or model codes
Catalog Number: MRDP50AXS, Lot Number: MPDX180, UDI Number: 884908149753 (updated 3/11/2021 added lot number: MPCY660)
510(k) numbers
["K132177"]
PMA numbers
not on file
Product code
LJT
Firm FEI
2518902
Firm city
Harleysville
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register