Siemens Medical Solutions USA, Inc: Siemens mMLC (ModuLeaf) safety bar and trolley label. Siemens Medical Solutions USA, Inc. Concord, CA. The intended use of the Siemens branded mMLC is an accessory used with the Linear Accelerator systems to deliver x-ray photon and electron radiation for the therapeutic treatment of cancer. This accessory shapes the radiation beam, per the leaf parameters, to minimize the amount of dose to outlying tissue.

Recall number
Z-1184-2012
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens mMLC (ModuLeaf) safety bar and trolley label. Siemens Medical Solutions USA, Inc. Concord, CA. The intended use of the Siemens branded mMLC is an accessory used with the Linear Accelerator systems to deliver x-ray photon and electron radiation for the therapeutic treatment of cancer. This accessory shapes the radiation beam, per the leaf parameters, to minimize the amount of dose to outlying tissue.
Status
Terminated
Initiated
2012-02-07
Posted
2012-03-07
Terminated
2012-08-14
Reason
This correction is being performed to update the safety knob of the Siemens brand mMLC (ModuLeaf) to conform to tie IEC 60601-1 safety standard.
Root cause
Nonconforming Material/Component
Action
The firm, SIEMENS, issued a "Customer Information" letter to its customers. The letter described the product, problem and actions to be taken. Siemens developed a new steel knob that complies with the latest standards; Siemens checked all ModuLeaf mMLC trolleys and if necessary refitted with the safety label and also arranged for an inspection and modification of the customers system- a Release of the safety update UI TH 026/11/S to distribute and hardware update commenced on February 7, 2007 by certified mail. If you have any questions, contact the Regulatory Specialist II at 925-602-8175.
Quantity
46 active devices
Distribution
Worldwide distribution: USA (nationwide) including states of: CO, IA, KY, NJ, OH and TX; and countries of: Algeria, Austria, Belgium, Brazil, China, Egypt, France, Germany, Hungary, India, Italy, Japan, Kuwait, Malaysia, Mexico, P.R., Russian Fed., Saudi Arabia, South Africa, Spain, Turkey, and United Kingdom.
Lot, serial or model codes
ModuLeaf, 110V/60Hz, part number 7334043; ModuLeaf, 220V/60Hz, part number 7334050.
510(k) numbers
["K030609"]
PMA numbers
not on file
Product code
IXI
Firm FEI
2937457
Firm city
Concord
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register