Beckman Coulter Inc: Beckman Coulter UniCel Dxl 800 Access Immunoassay System; Part Number: 973100 With: UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733), for in vitro diagnostic use.

Recall number
Z-1184-2008
Recalling firm
Beckman Coulter Inc
Product
Beckman Coulter UniCel Dxl 800 Access Immunoassay System; Part Number: 973100 With: UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733), for in vitro diagnostic use.
Status
Terminated
Initiated
2007-10-08
Posted
2008-08-20
Terminated
2012-01-11
Reason
Incorrect results: There is an increased variance in assay performance when the UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733) have been implemented and there are room fluctuations after assay calibration between the extremes of the 18C and 32C operating range.
Root cause
Device Design
Action
Beckman Coulter mailed a Product Corrective Action (PCA) letter the week of October 8, 2007 to the customers who have the affected UniCel Dxl 800 Access lmmunoassay systems advising that there is an increased variance in assay performance when the UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733) have been implemented and there are room fluctuations after assay calibration between the extremes of the 18C and 32C operating range. The letter further advises that Beckman Coulter will contact the customer to arrange a date to remove the modificatlons but until the modification is completed, the system should be operated in a room temperature of 20C to 30C. This can be acomplished by monitoring and recording room tamperature in the area of the instrument. Customers are also asked to review historical quality control records to ensure assay performance is within their laboratory specification and that patient results should be reviewed in context of other diagnostics tests and the clinical presentations of the patient. A response form is included.
Quantity
111 units in the US; 9 units in Canada
Distribution
Nationwide and Canada
Lot, serial or model codes
Part Number 973100; Serial Numbers: 600469, 601097, 601098, 601144, 601182, 601160, 601188, 601071, 601117, 601114, 601137, 601059, 601039, 601099, 601088, 601176, 601184, 601178, 601025, 601186, 601110, 601063, 601002, 601029, 601133, 601106, 601215, 601159, 601077, 601202, 601206, 601079, 601051, 601179, 601130, 601058, 601031, 601212, 601173, 601028, 601078, 601094, 601210, 601189, 601228, 601043, 601092, 601171, 601026, 601122, 601208, 601103, 601089, 601108, 601183, 600933 600971, 600919, 600910, 600904, 600935, 601005, 600977, 600936, 600993, 600965, 601011, 600957, 600918, 601015, 600992, 600905, 600914, 600947, 601016, 601013, 600981, 600903, 600949, 601012, 600983, 600970, 600912, 601004, 600915, 600907, 600987, 600939, 600996, 601007, 600899, 601006, 600938, 600952, 600988, 601020, 600928, 600937, 600972, 955056, 965059, 954609, 949471, 954490, 501442, and 499300.
510(k) numbers
["K023764"]
PMA numbers
not on file
Product code
JJE
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register