Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc: Access Power Supply Assembly Sled Part Number: A48196/6845C; Component of: Access and Access 2 Immunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC 600i System, for in vitro diagnostic use.

Recall number
Z-1183-2008
Recalling firm
Beckman Coulter Inc
Product
Access Power Supply Assembly Sled Part Number: A48196/6845C; Component of: Access and Access 2 Immunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC 600i System, for in vitro diagnostic use.
Status
Terminated
Initiated
2007-10-17
Posted
2008-08-20
Terminated
2012-04-27
Reason
Electrical grounding failure: During manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering.
Root cause
Process control
Action
Beckman Coulter mailed a Product Corrective Action (PCA) letter on October 17, 2007 to all Access and Access 2 lmmunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC 600i System customers. The customers were informed that during manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering. The letter referenced a proactive initiative to inspect the power supply sled assemblies on all instruments referenced. A response form was included.
Quantity
3055 instruments in the US; 192 instruments in Canada
Distribution
Nationwide Distribution and Canada
Lot, serial or model codes
Part Number A48196/6845C; Instrument part numbers 81600 and  81600N
510(k) numbers
["K023049","K060256","K922823"]
PMA numbers
not on file
Product code
JJE
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register