Skip to the content

FDA Medical Device Recalls (openFDA)

Alphatec Spine, Inc.: IdentiTi ALIF Standalone Interbody System: Part Number/Description 132-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm 132-10-85-300 / IdentiTi ALIF SA Graft Bolt, 08.5 x 30 mm 132-10-85-350 / IdentiTi ALIF SA Graft Bolt, 08.5 x 35 mm 132-10-90-250 / IdentiTi ALIF SA Graft Bolt, 09 x 25 mm 132-10-90-300 / IdentiTi ALIF SA Graft Bolt, 09 x 30 mm 132-10-90-350 / IdentiTi ALIF SA Graft Bolt, 09 x 35 mm 232-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm 232-10-85-300 / IdentiTi ALIF SA Graft Bolt, 08.5 x 30 mm 232-10-85-350 / IdentiTi ALIF SA Graft Bolt, 08.5 x 35 mm 232-11-90-250 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 25 mm 232-11-90-300 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 30 mm 232-11-90-350 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 35 mm The IdentiTi ALIF Standalone Interbody System is an integrated intervertebral body fusion device for use in anterior lumbar interbody fusion (ALIF) procedures.

Recall number
Z-1182-2024
Recalling firm
Alphatec Spine, Inc.
Product
IdentiTi ALIF Standalone Interbody System: Part Number/Description 132-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm 132-10-85-300 / IdentiTi ALIF SA Graft Bolt, 08.5 x 30 mm 132-10-85-350 / IdentiTi ALIF SA Graft Bolt, 08.5 x 35 mm 132-10-90-250 / IdentiTi ALIF SA Graft Bolt, 09 x 25 mm 132-10-90-300 / IdentiTi ALIF SA Graft Bolt, 09 x 30 mm 132-10-90-350 / IdentiTi ALIF SA Graft Bolt, 09 x 35 mm 232-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm 232-10-85-300 / IdentiTi ALIF SA Graft Bolt, 08.5 x 30 mm 232-10-85-350 / IdentiTi ALIF SA Graft Bolt, 08.5 x 35 mm 232-11-90-250 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 25 mm 232-11-90-300 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 30 mm 232-11-90-350 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 35 mm The IdentiTi ALIF Standalone Interbody System is an integrated intervertebral body fusion device for use in anterior lumbar interbody fusion (ALIF) procedures.
Status
Open, Classified
Initiated
2024-01-26
Posted
2024-02-23
Terminated
not on file
Reason
Due to reports of intraoperative graft bolt implantation breakages.
Root cause
Device Design
Action
On 01/26/2024, the firm emailed an "URGENT-MEDICAL DEVICE RECALL NOTIFICATION" Letter to customer informing them that, The Surgical Technique Guide, LIT-85141, and Instructions for Use, INS-130, were revised to require hole preparation prior to insertion of an IdentiTi SA Graft Bolt, as well as to add a new caution statement, to confirm with fluoroscopy that the graft bolt insertion angle is as close as possible to a 40-degree trajectory. The revised labeling mitigates potential fracturing of the Graft Bolt intra-operatively or post-operatively. Customer are instructed to: Please share this notice with all of those who need to be aware within your organization or any organization where the affected product has been transferred. - Inform doctors and surgical staff of the labeling (surgical technique and instructions for use) - Please fill out the last page of this letter to confirm that you have read this notification and will complete the necessary actions as described in this notification. - If you have any questions/concerns or require assistance with the correction, please contact by e-mailing QMSCompliance@atecspine.com or calling 1-800-922-1356.
Quantity
12,777 devices
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IN, KS, KY, MD, MI, MN, MS, NC, NH, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, WA and WI. The country of New Zealand.
Lot, serial or model codes
Part Number / Product Description / UDI-DI Code / Lot Number 132-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm UDI-DI code: 00190376299716 Lot Numbers: 8746709, 8746724, TO00212 132-10-85-300 / IdentiTi ALIF SA Graft Bolt, 08.5 x 30 mm UDI-DI code: 00190376299730 Lot Numbers: 8746710, 8746725, TO00213 132-10-85-350 / IdentiTi ALIF SA Graft Bolt, 08.5 x 35 mm UDI-DI code: 00190376299754 Lot Numbers: 8746711, TO00218 132-10-90-250 / IdentiTi ALIF SA Graft Bolt, 09 x 25 mm UDI-DI code: 00190376299792 Lot Numbers: 8746712 132-10-90-300 / IdentiTi ALIF SA Graft Bolt, 09 x 30 mm UDI-DI code: 00190376299815 Lot Numbers: 6746713 132-10-90-350 / IdentiTi ALIF SA Graft Bolt, 09 x 35 mm UDI-DI code: 00190376299839 Lot Numbers: 8746714 232-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm UDI-DI code: 00190376369976 Lot Numbers: TO00373, TO00670, TO00694, TO00834, TO01038, TO01404, TO01514, TO01572, TO01573, TO01696, TO01974, TO01979, TO01982, TO02297, TO02393, TO03218, TO03465 232-10-85-300 / IdentiTi ALIF SA Graft Bolt, 08.5 x 30 mm UDI-DI code: 00190376369990 Lot Numbers: TO00376, TO00671, TO00695, TO00710, TO00835, TO01039, TO01405, TO01515, TO01574, TO01575, TO01697, TO01975, TO01980, TO01983, TO02298, TO02394, TO03219, TO03466 232-10-85-350 / IdentiTi ALIF SA Graft Bolt, 08.5 x 35 mm UDI-DI code: 00190376370019 Lot Numbers: TO00380, TO00672, TO00696, TO00836, TO01576, TO01577, TO02163, TO02395 232-11-90-250 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 25 mm UDI-DI code: 00190376370293 Lot Numbers: TO00381, TO00673, TO00697, TO01277, TO01578, TO02396 232-11-90-300 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 30 mm UDI-DI code: 00190376370316 Lot Numbers: TO00382, TO00674, TO00698, TO01040, TO01406, TO02164, TO02397, TO03220 232-11-90-350 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 35 mm UDI-DI code: 00190376370330 Lot Numbers: TO00383, TO00699, TO01407, TO02398
510(k) numbers
["K232097"]
PMA numbers
not on file
Product code
OVD
Firm FEI
2027467
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register