Skip to the content

FDA Medical Device Recalls (openFDA)

Covidien, LP: Custom Suture Pack Product Description: US1638 CUSTOM SUTURE PACK US1756 CUSTOM SUTURE PACK

Recall number
Z-1182-2023
Recalling firm
Covidien, LP
Product
Custom Suture Pack Product Description: US1638 CUSTOM SUTURE PACK US1756 CUSTOM SUTURE PACK
Status
Open, Classified
Initiated
2022-12-02
Posted
2023-02-25
Terminated
not on file
Reason
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
Root cause
Packaging process control
Action
Medtronic issued an "URGENT: MEDICAL DEVICE RECALL" letter to US consignees on 01-Dec-2022 via UPS 2-day delivery. OUS countries, notifications are handled via a regionally approved method. Letter states reason for recall, health risk and actions to take: Actions: " Identify and quarantine all unused and non-expired affected non-absorbable sutures. " Please note the affected product may be located within Custom Suture Packs. Please reference Custom Suture Packs to identify affected product. " Please complete the enclosed Customer Confirmation Form (even if your account no longer has affected product on hand) and email to rs.gmbfcamitg@medtronic.com.> " Return all unused and non-expired affected product in your inventory to Medtronic as indicated in the Shipping and Return Instructions below. " Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected product has been transferred or distributed. Shipping and Return Instructions: " Product purchased directly from Medtronic please contact rs.covidienfeedbackcustomerservice@medtronic.com for Return Good Authorization (RGA). Your local Medtronic Representative can assist you as necessary in initiating the return of this product. " Credit or Replacement for the returned affected product will be issued based on the RGA number. " Return all unused and non-expired affected product(s) in your inventory to: Medtronic, Attn: Field Returns Dept. 195 McDermott Road, North Haven, CT 06473 USA. " If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. If you have any questions regarding this communication, please contact your Medtronic Representative or Customer Service at 800-962-9888.
Quantity
9606
Distribution
US Nationwide Distribution and OUS Foreign distribution to countries of: Albania Argentina Armenia Australia Austria Azerbaijan Belarus Belgium Bosnia And Herzegovina Brazil Bulgaria Cambodia Cameroon Canada Canary Islands Chile China Colombia Croatia Czech Republic Denmark Egypt Finland France Georgia Germany Ghana Greece Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy JAPAN Jordan Kazakhstan Kenya Korea, Republic Of Kosovo Kuwait Latvia Lebanon Lithuania Macao Malaysia Malta Mauritius Mexico Moldova, Republic Of Morocco Myanmar Namibia Netherlands New Zealand North Macedonia Norway Pakistan Panama Paraguay Philippines Poland Portugal Reunion Romania Russian Federation Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Syrian Arab Republic Taiwan Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Viet Nam Zimbabwe
Lot, serial or model codes
GTIN: N/A Model Number Lot # US1638: A1K0289Y, A1K0291Y, US1638 A1K0283Y ; US1756: A2E0454Y
510(k) numbers
not on file
PMA numbers
not on file
Product code
GAT
Firm FEI
1219930
Firm city
North Haven
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register